Hemodialysis

Can Bioimpedance Analysis Be Used to Estimate Absolute Blood Volume in Hemodialysis Patients?

Hemodial IntPublished Method comparison study

Authors' conclusion

In almost half of patients, blood volume can be inferred from bioimpedance data with sufficient certainty. But the greater the deviation from the physiological blood to extracellular volume ratio of 1-3, the more the calculated blood volumes differ from the measured values. For this reason, bioimpedance data should not be used uncritically to set the ultrafiltration.

Study context

Population
HD patients
Comparison
Blood volume calculated from bioimpedance extracellular volume divided by 3 vs blood volume measured by indicator dilution
Outcome
Agreement of blood volume calculated from bioimpedance extracellular volume with measured blood volume
Follow-up
Immediately before treatment
Methodological detailsindicator dilution, bioimpedance spectroscopy
  • Indicator dilution
  • Bioimpedance spectroscopy

Main finding

As reported in the abstract

Blood volume
Intra-individual discrepancies, range
−1.409 to 1.450 L
Median absolute deviation
382 mL
Median absolute deviation per kg
6.2 mL/kg

Further findings

Correlation of measured–calculated difference with blood to extracellular volume ratio

R
−0.98
P
P < 0.001

Original publication

Can Bioimpedance Analysis Be Used to Estimate Absolute Blood Volume in Hemodialysis Patients? Hemodial Int. Published 5 March 2025. doi:10.1111/hdi.13217