Concordance of a Laboratory-Data Integrated Fluid Targeting Algorithm with Bioimpedance-Derived Volume Measures in Maintenance Hemodialysis
Main finding · Extracellular waterResult — go to the resultMean difference +0.22 LAuthors' conclusion
The laboratory-based algorithm approximated Body Composition Monitor outputs at the cohort level; however, proportional bias and wide individual limits, particularly for intracellular water, preclude interchangeability. These findings support method concordance but do not constitute independent physiological or outcome validation. Clinical actionability and outcome benefit remain untested.
Study context
- Population
- Adults receiving stable maintenance hemodialysis for at least 3 months
- Participants
- 61 adults
- Exposure
- Laboratory-Data Integrated Fluid Targeting algorithm vs Fresenius Body Composition Monitor
- Outcome
- Extracellular water
Methodological detailspaired assessment during one dialysis session, Pearson correlation, Deming regression, and Bland-Altman analysis, exploratory multivariable models
- Paired assessment during one dialysis session
- Pearson correlation, Deming regression, and Bland-Altman analysis
- Exploratory multivariable models
Main finding
Extracellular water
- Correlation coefficient
- 0.87
Key notes
- 95% limits of agreement from −4.1 to +4.6 L
Observational data: associations, not effects.
Further findings
| Outcome | Estimate |
|---|---|
| Intracellular waterLimits of agreement from −3.5 to +8.9 L | mean difference +2.7 L |
| Estimated euvolemic weightLimits of agreement, −4.0 to +2.6 kg · partly reflected the shared body-weight component | mean difference −0.67 kg |
- Intracellular water — Correlation coefficient
- 0.80
- Estimated euvolemic weight — Correlation coefficient
- 1.0 after rounding
Determinants of method differences
- Older age was associated with a smaller extracellular-water difference
- Sex, dialysis vintage, diabetes, heart failure, and ischemic heart disease were not independently associated with method differences
Original publication
Concordance of a Laboratory-Data Integrated Fluid Targeting Algorithm with Bioimpedance-Derived Volume Measures in Maintenance Hemodialysis. Kidney360. Published 16 November 2026. doi:10. 34067/ kid. 0000001468