Hemodialysis

A Pilot Study on the Safety and Adequacy of a Novel Ecofriendly Hemodialysis Prescription–Green Nephrology

Kidney Int RepPublished Nonrandomised crossover pilot trial

Main finding · Adequate dialysis (spKt/V ≥1.2)Result — go to the resultOdds ratio 1.07 95% CI 0.75–1.52

Authors' conclusion

Increasing dialysate temperature while reducing dialysate flow rate (QD) can be used as a water conservation strategy without compromising the adequacy and safety of dialysis in young and hemodynamically stable patients. Reducing the QD from 500 ml/min to 300 ml/min reduces water consumption by 40%.

Study context

Population
Patients with ESKD on maintenance hemodialysis (HD)863 HD sessions
Participants
30 patients
Comparison
High temperature with low flow dialysate (HTLF) vs normal temperature with normal flow dialysate (NTNF), 1 month each
Outcome
Delivery of adequate dialysis, defined by spKt/V ≥1.2
Follow-up
End of each dialysis session
Methodological detailssingle-center, nonrandomized, open-label, cross-over
  • Single-center
  • Nonrandomized
  • Open-label
  • Cross-over

Main finding

Adequate dialysis (spKt/V ≥1.2) · end of each session

Odds ratio

1.07

95% CI 0.75–1.52

P = 0.45

Events by group

NTNFvs HTLF
OR 1.07 (95% CI 0.75–1.52); P = 0.45

As reported in the abstract

Further findings

Mean spKt/V and urea reduction ratio (URR) not significantly different

Safety

Clinically evident hemodynamic instability — HTNF
1 dialysis session events
  • Clinically evident hemodynamic instability: occurred in only 1 session, in HTNF

Original publication

A Pilot Study on the Safety and Adequacy of a Novel Ecofriendly Hemodialysis Prescription–Green Nephrology. Kidney Int Rep. Published 13 February 2024. doi:10.1016/j.ekir.2024.02.014

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.