Hemodynamic Monitoring During Hemodialysis Using Bioimpedance: A Comparison of Changes in Resistance Between Different Body Segments
Authors' conclusion
These findings suggest that the thorax is the most suitable region for segmental bioimpedance measurements to assess hemodynamic parameters. Thoracic bioimpedance may innovate the hemodynamic monitoring of hemodialysis patients.
Study context
- Population
- Patients on HD
- Participants
- 15 patients
- Exposure
- Bioimpedance resistance of whole body vs lower leg vs upper arm vs thorax
- Outcome
- Which body segment is best targeted to measure bioimpedance for hemodynamic monitoring: relationship of relative resistance with changes in volume and SBP
Methodological detailsserial bioimpedance measurements, two HD sessions per patient
- Serial bioimpedance measurements
- Two HD sessions per patient
Main finding
As reported in the abstract
Correlation of relative resistance with changes in SBP
| Group | Value | P |
|---|---|---|
| Thorax | correlation −0.33 (CI −0.432 to −0.219) | p < 0.001 |
| Whole body | correlation −0.154 (CI −0.269 to −0.036) | p = 0.01 |
Value
Thoraxcorrelation −0.33 (CI −0.432 to −0.219)
Whole bodycorrelation −0.154 (CI −0.269 to −0.036)
P
Thoraxp < 0.001
Whole bodyp = 0.01
- Multiple regression: thoracic resistance best predictor of changes in systolic blood pressure (SBP), β
- −0.261 [−0.353 to −0.126], p < 0.001
Key notes
- Highest correlation in the thoracic segment, lowest for whole body
Observational data: associations, not effects.
Further findings
| Outcome | Whole body | Lower leg | Upper | Thorax |
|---|---|---|---|---|
| Correlation of UF volume with relative resistance | 0.697 (CI 0.63–0.754) | 0.670 (CI 0.598–0.731) | 0.538 (CI 0.447–0.618) | 0.537 (CI 0.446–0.617) |
| P | p < 0.001 | p < 0.001 | p < 0.001 | p < 0.001 |
| Reproducibility of pre- to postdialysis change in relative resistance between sessions (intraclass correlation coefficient) | — not stated in the abstract | — not stated in the abstract | 0.023 | 0.728 |
— not stated in the abstract
Correlation of UF volume with relative resistance
- Comparable for upper arm and thorax; highest for whole body and lower leg
Reproducibility of pre- to postdialysis change in relative resistance between sessions (intraclass correlation coefficient)
- Lowest for upper arm, highest for thoracic resistance
Original publication
Hemodynamic Monitoring During Hemodialysis Using Bioimpedance: A Comparison of Changes in Resistance Between Different Body Segments. Hemodial Int. Published 28 April 2025. doi:10. 1111/ hdi. 13245