Accuracy of Bioimpedance Spectroscopy in the Detection of Hydration Changes in Patients on Hemodialysis
Authors' conclusion
The accuracy of BCM volume change estimates is highly variable and below requirements of daily practice. Improvement may be achieved by a switch to an end-of-HD measurement.
Study context
- Population
- Patients on hemodialysis (HD)24 for precision, 45 for accuracy
- Comparison
- Body composition monitor (BCM) ΔOH during HD vs ultrafiltration (UF) volume
- Outcome
- Accuracy: change in OH (ΔOH) during HD vs UF volume; acceptable if volume error in individual patients ≤0.5 L
Performance
As reported in the abstract
Accuracy of OH change estimate (ΔOH vs UF volume)
- Correlation between UF volume and ΔOH moderate: slope
- 0.66
- R2
- 0.44
- P
- P < .001
- Patients with volume error >0.5 L
- 44%
Key notes
- Underestimation of UF volume, in particular for high UF volumes
- Accuracy at individual level highly variable
More notes
- Accuracy acceptable if individual volume error ≤0.5 L
Further findings
Analytical precision of the overhydration (OH) estimate
- Analytical precision of OH estimate
- 1.0 ± 0.4%
- 6 consecutive measurements in 24 patients on HD
Accuracy over the course of HD
L
- Total error range, first hour of HD
- 2.3
- Total error range, final hour of HD
- 1.1
- Accuracy improved over the course of HD