Hemodialysis

Accuracy of Bioimpedance Spectroscopy in the Detection of Hydration Changes in Patients on Hemodialysis

J Ren NutrPublished Diagnostic accuracy study

Authors' conclusion

The accuracy of BCM volume change estimates is highly variable and below requirements of daily practice. Improvement may be achieved by a switch to an end-of-HD measurement.

Study context

Population
Patients on hemodialysis (HD)24 for precision, 45 for accuracy
Comparison
Body composition monitor (BCM) ΔOH during HD vs ultrafiltration (UF) volume
Outcome
Accuracy: change in OH (ΔOH) during HD vs UF volume; acceptable if volume error in individual patients ≤0.5 L

Performance

As reported in the abstract

Accuracy of OH change estimate (ΔOH vs UF volume)
Correlation between UF volume and ΔOH moderate: slope
0.66
R2
0.44
P
P < .001
Patients with volume error >0.5 L
44%

Further findings

Analytical precision of the overhydration (OH) estimate

Analytical precision of OH estimate
1.0 ± 0.4%
  • 6 consecutive measurements in 24 patients on HD

Accuracy over the course of HD

L

Total error range, first hour of HD
2.3
Total error range, final hour of HD
1.1
  • Accuracy improved over the course of HD

Original publication

Accuracy of Bioimpedance Spectroscopy in the Detection of Hydration Changes in Patients on Hemodialysis. J Ren Nutr. Published 23 December 2021. doi:10.1053/j.jrn.2021.11.004