Hemodialysis

Catheter Exchange With Elongation of Tunnel (CEET) Procedure—A Novel Technique for Cuff Extrusion of Tunneled Dialysis Catheter: Surgical Experience and Early Outcomes

Semin DialPublished Retrospective study

Authors' conclusion

CEET procedure is a successful alternative technique for correction of cuff extrusion of TDC with good success rate. Premature removal of anchor sutures was associated with early cuff extrusion, whereas short tunnel length was associated with late cuff extrusion.

Study context

Population
HD patients with tunneled dialysis catheter (TDC) with partial or complete cuff extrusion, excluding complete catheter dislodgement, tunnel infection or any catheter-related infection
Participants
11 patients
Outcome
CEET procedure success (desired catheter tip position and good blood flow)
Methodological detailsfluoroscopy-guided
  • Fluoroscopy-guided

Main finding

As reported in the abstract

CEET procedure success (desired catheter tip position and good blood flow)

Successful CEET procedure

10 cases

Success rate

90.1%

Further findings

Case characteristics

Previous cuff extrusion
three (27.2%)
Partial cuff extrusion
four (36.4%)
Short tunnel length
seven cases (63.6%)

Factors associated with early vs late (≥1 month) cuff extrusion

Short tunnel length associated with late extrusion
P = 0.05
Premature removal of TDC anchor sutures associated with early cuff extrusion
P = 0.04

Original publication

Catheter Exchange With Elongation of Tunnel (CEET) Procedure—A Novel Technique for Cuff Extrusion of Tunneled Dialysis Catheter: Surgical Experience and Early Outcomes. Semin Dial. Published 24 October 2024. doi:10.1111/sdi.13230

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.