Hemodialysis

The Emboless® venous chamber efficiently reduces air bubbles: a randomized study of chronic hemodialysis patients

Clin Kidney JPublished Randomised crossover trial

Authors' conclusion

Fewer MBs and subsequently fewer microemboli entered the bloodline of the patient using the Emboless-VC compared to the F5008-VC venous chamber during HD and during HDF. Venous chambers with a high removal rate of MBs will reduce the extent of air emboli.

Study context

Population
Chronic hemodialysis (HD) patients80 sessions
Participants
20 patients
Comparison
Emboless venous chamber vs Fresenius 5008 venous chamber
Primary endpoint
Elimination of air microbubbles (MBs) in the return bloodline during HD (% MB reduction, inlet vs outlet of venous chamber, ultrasound, 20-500 µm)
Methodological detailscross-over, randomized order
  • Cross-over
  • Randomized order

Primary result

As reported in the abstract

Median reduction of MBs at outlet during HD

Emboless-VC

76%

median

F5008-VC

39%

median

P < 0.001

Further findings

Reduction of MBs during HDF

P < 0.001

  • Reduction of microbubbles (MBs) during hemodiafiltrations (HDF): 32 of the 80 sessions, HDF after initial HD

Original publication

The Emboless® venous chamber efficiently reduces air bubbles: a randomized study of chronic hemodialysis patients. Clin Kidney J. Published 24 October 2024. doi:10.1093/ckj/sfae323

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.