The Emboless® venous chamber efficiently reduces air bubbles: a randomized study of chronic hemodialysis patients
Authors' conclusion
Fewer MBs and subsequently fewer microemboli entered the bloodline of the patient using the Emboless-VC compared to the F5008-VC venous chamber during HD and during HDF. Venous chambers with a high removal rate of MBs will reduce the extent of air emboli.
Study context
- Population
- Chronic hemodialysis (HD) patients80 sessions
- Participants
- 20 patients
- Comparison
- Emboless venous chamber vs Fresenius 5008 venous chamber
- Primary endpoint
- Elimination of air microbubbles (MBs) in the return bloodline during HD (% MB reduction, inlet vs outlet of venous chamber, ultrasound, 20-500 µm)
Methodological detailscross-over, randomized order
- Cross-over
- Randomized order
Primary result
As reported in the abstract
Median reduction of MBs at outlet during HD
Emboless-VC
76%
median
F5008-VC
39%
median
P < 0.001
Key notes
- Wilcoxon pairwise test
Further findings
Reduction of MBs during HDF
P < 0.001
- Reduction of microbubbles (MBs) during hemodiafiltrations (HDF): 32 of the 80 sessions, HDF after initial HD
Original publication
The Emboless® venous chamber efficiently reduces air bubbles: a randomized study of chronic hemodialysis patients. Clin Kidney J. Published 24 October 2024. doi:10. 1093/ ckj/ sfae323