Automated measurement of intraperitoneal pressure versus patient fill-volume
Authors' conclusion
In summary, this study indicates that a transducer-based IPP measurement device could serve as an alternative to the Durand method.
Study context
- Population
- Patients with kidney failure on automated peritoneal dialysis (APD) therapy, in-clinic APD70% female
- Participants
- 10 patients
- Comparison
- Intraperitoneal pressure (IPP) by in-line pressure transducer vs Durand method
- Outcome
- Agreement between IPP measured by in-line pressure transducer and by Durand method
Methodological detailsexploratory, descriptive, early feasibility, Median age 47.0 years; median BMI 28.1 kg/m²
- Exploratory
- Descriptive
- Early feasibility
- Median age 47.0 years; median BMI 28.1 kg/m²
Main finding
As reported in the abstract
Agreement of transducer IPP with manual IPP
Moderate agreement; most differences within
±3 cmH2O
Further findings
- Patients with stronger correlations (ρ > 0.7)
- 6
- Patients with weaker correlations (ρ < 0.5)
- 4
- Overall, IPP did not correlate with fill-volume
- Per-patient level
Safety
- No safety concerns identified
Original publication
Automated measurement of intraperitoneal pressure versus patient fill-volume. Perit Dial Int. Published 7 April 2026. doi:10. 1177/ 08968608261435113