Automated measurement of intraperitoneal pressure versus patient fill-volume

Perit Dial IntPublished Pilot study

Authors' conclusion

In summary, this study indicates that a transducer-based IPP measurement device could serve as an alternative to the Durand method.

Study context

Population
Patients with kidney failure on automated peritoneal dialysis (APD) therapy, in-clinic APD70% female
Participants
10 patients
Comparison
Intraperitoneal pressure (IPP) by in-line pressure transducer vs Durand method
Outcome
Agreement between IPP measured by in-line pressure transducer and by Durand method
Methodological detailsexploratory, descriptive, early feasibility, Median age 47.0 years; median BMI 28.1 kg/m²
  • Exploratory
  • Descriptive
  • Early feasibility
  • Median age 47.0 years; median BMI 28.1 kg/m²

Main finding

As reported in the abstract

Agreement of transducer IPP with manual IPP

Moderate agreement; most differences within

±3 cmH2O

Further findings

Patients with stronger correlations (ρ > 0.7)
6
Patients with weaker correlations (ρ < 0.5)
4
Overall, IPP did not correlate with fill-volume
  • Per-patient level

Safety

No safety concerns identified

Original publication

Automated measurement of intraperitoneal pressure versus patient fill-volume. Perit Dial Int. Published 7 April 2026. doi:10.1177/08968608261435113