Prognostic Value of Remote Monitoring Alarms for Technique Failure in Automated Peritoneal Dialysis
Main finding · Death-censored technique failure (permanent transfer to HD)Result — go to the resultAdjusted odds ratio 1.48 95% CI 1.13–2.04Authors' conclusion
An RPM alarm for initial drain variance >50% emerged as an early marker of technique failure in APD. Incorporating alarm-pattern analysis into routine RPM review may facilitate targeted surveillance and timely intervention for high-risk patients.
Study context
- Population
- Adult automated peritoneal dialysis (APD) patients managed with the Sharesource remote patient monitoring (RPM) platform at National Cheng Kung University Hospital
- Participants
- 67 patients272 alarm episodes
- Exposure
- Tertiles of RPM alarm frequency; red-flag alarm types
- Outcome
- Death-censored technique failure, defined as permanent transfer to HD
- Follow-up
- 3 years
Methodological detailssingle-centre, logistic regression, Fine and Gray competing-risk and GEE sensitivity analyses
- Single-centre
- Logistic regression
- Fine and Gray competing-risk and GEE sensitivity analyses
Main finding
As reported in the abstract
Death-censored technique failure (permanent transfer to HD)
Adjusted odds ratio
1.48
95% CI 1.13–2.04
P = 0.006
Initial drain variance > 50%
aOR 1.48 (95% CI 1.13–2.04); P = 0.006
- Technique failure, all patients
- 8 (11.9%)
- Technique failure across alarm-frequency tertiles
- 4.5%, 9.1%, and 21.7%; p for trend = 0.074
Key notes
- Multivariable analysis
Observational data: associations, not effects.
Further findings
Technique failure, sensitivity analyses
- Fine and Gray subdistribution HR
- 1.55 (95% CI: 1.07-2.25)
- GEE OR
- 2.58 (95% CI: 1.00-6.62)
- Results consistent in sensitivity analyses
Other alarm types (lost therapy volume >10%, lost treatment time >30 min)
- The other two alarm types were not significantly associated with outcomes
Original publication
Prognostic Value of Remote Monitoring Alarms for Technique Failure in Automated Peritoneal Dialysis. Blood Purif. Published 2 April 2026. doi:10. 1159/ 000551828
Composed by AI from the published abstract and checked against it. How Briefs are made →