Electrolytes, acid–base & stones
Validation of Noninvasive Detection of Hyperkalemia by Artificial Intelligence–Enhanced Electrocardiography in High Acuity Settings
Authors' conclusion
The AI-ECG algorithm demonstrated a high NPV, suggesting that it is useful for ruling out hyperkalemia, but a low PPV, suggesting that it is insufficient for treating hyperkalemia.
Study context
- Population
- Emergency department (ED) and mixed intensive care unit (ICU) patients with laboratory potassium and 12-lead electrocardiogram (ECG) obtained within 4 hours of each other2 cohorts
- Participants
- 40,128 vs 2,636
- Comparison
- Artificial intelligence (AI)-enhanced ECG vs laboratory potassium for detection of hyperkalemia in ED and ICU cohorts
- EDn = 40,128
- ICUn = 2,636
- Outcome
- AI-ECG detection of hyperkalemia (potassium >6.0 mEq/L)
Methodological detailsvalidation of a previously developed AI-ECG algorithm, leads 1 and 2 of 12-lead ECG, potassium and ECG pairs within 4 hours, 2 cohorts analysed separately
- Validation of a previously developed AI-ECG algorithm
- Leads 1 and 2 of 12-lead ECG
- Potassium and ECG pairs within 4 hours
- 2 cohorts analysed separately
Performance
As reported in the abstract
Detection of hyperkalemia by AI-ECG
| Test or model | AUC | Sensitivity | Specificity | PPV | NPV | LR+ |
|---|---|---|---|---|---|---|
| Emergency departmentED | 0.88 | 80% | 80% | 3% | 99.8% | 4.0 |
| Intensive care unitICU | 0.88 | 82% | 82% | 14% | 99% | 4.6 |
AUC
Emergency departmentED0.88
Intensive care unitICU0.88
Sensitivity
Emergency departmentED80%
Intensive care unitICU82%
Specificity
Emergency departmentED80%
Intensive care unitICU82%
PPV
Emergency departmentED3%
Intensive care unitICU14%
NPV
Emergency departmentED99.8%
Intensive care unitICU99%
LR+
Emergency departmentED4.0
Intensive care unitICU4.6
Further findings
Detection of hyperkalemia by AI-ECG, low-eGFR (<30 ml/min) subanalysis
AUC ED 0.83 · ICU 0.85
Sensitivity
Specificity
PPV
NPV
LR+ ED 2.2 · ICU 2.7