Use of a sampler to collect spent dialysate in hemodialysis: a technical note
Authors' conclusion
The Theraflux® sampler demonstrated satisfactory reliability for partial spent dialysate collection during hemodiafiltration sessions. Its automated sampling system substantially reduces the logistical burden associated with total dialysate collection and may facilitate future studies assessing toxin mass transfer and albumin loss in routine clinical practice.
Study context
- Population
- Chronic dialysis patients undergoing hemodiafiltration in a French dialysis center16 analysed
- Participants
- 20 patients
- Exposure
- Theraflux® sampler-estimated vs dialysis machine-calculated total dialysate volume
- Outcome
- Agreement between sampler-estimated and machine-calculated total dialysate volume
Methodological detailssingle-centre, repeated-measures Bland-Altman analysis, French dialysis center, November 2021 to February 2022
- Single-centre
- Repeated-measures Bland-Altman analysis
- French dialysis center, November 2021 to February 2022
Main finding
As reported in the abstract
Difference between machine-calculated and sampler-reported dialysate volume
Key notes
- IQR 4.3-14.8; median discrepancy 7.4% (IQR 4.5-11.6%)
Observational data: associations, not effects.
Further findings
Median
| Outcome | Theraflux® sampler | Dialysis machine |
|---|---|---|
| Fraction of total dialysate collected by sampler | 0.96% (IQR 0.96–0.97%) | 1.02% (IQR 1.00–1.07%) |
- Fraction of total dialysate collected by sampler: denominator: total dialysate per machine calculation vs per sampler-derived values
Outlier sessions
- Outlier sessions
- 12 (9.4%)
- Excluding outliers improved agreement without reducing the average bias
Original publication
Use of a sampler to collect spent dialysate in hemodialysis: a technical note. J Nephrol. Published 6 August 2026. doi:10. 1093/ joneph/ aajag217