Detection of Hemodialysis Venous Needle Dislodgment Using Venous Access Pressure Measurements: A Simulation Study

Kidney360Published Simulation study with cross-sectional retrospective study

Authors' conclusion

Patients can experience changes in VP during hemodialysis which may not trigger a machine alarm in the case of a VND. Our simulations showed that current dialysis machine alarm systems may not compensate for upward drift in VP, and improved algorithms for detecting needle dislodgment during hemodialysis are needed.

Study context

Population
Patients on HD with an AVF (Sept 1–Nov 1, 2016); in vitro venous needle dislodgment (VND) simulations43,390 venous pressure (VP) readings
Participants
147 patients
Exposure
Improved alarm algorithm vs dialysis machine alarm for detection of VND
Outcome
Alarm activation in VND simulations
Methodological detailsin vitro sham dialysis circuit (Fresenius 2008K), blood substitute
  • In vitro sham dialysis circuit (Fresenius 2008K)
  • Blood substitute

Main finding

As reported in the abstract

Alarm activation in VND simulations
VND simulations
19

Observational data: associations, not effects.

Further findings

Upward drift in VP during HD

VP readings with upward drift
16,594 (38%)
Range of upward drift
20-79 mmHg
Increase in VP, mean±SD
11±18 mm Hg
Within, minutes
20±14

Original publication

Detection of Hemodialysis Venous Needle Dislodgment Using Venous Access Pressure Measurements: A Simulation Study. Kidney360. Published 24 March 2023. doi:10.34067/kid.0000000000000093

Composed by AI from the published abstract and checked against it. How Briefs are made →