Detection of Hemodialysis Venous Needle Dislodgment Using Venous Access Pressure Measurements: A Simulation Study
Authors' conclusion
Patients can experience changes in VP during hemodialysis which may not trigger a machine alarm in the case of a VND. Our simulations showed that current dialysis machine alarm systems may not compensate for upward drift in VP, and improved algorithms for detecting needle dislodgment during hemodialysis are needed.
Study context
- Population
- Patients on HD with an AVF (Sept 1–Nov 1, 2016); in vitro venous needle dislodgment (VND) simulations43,390 venous pressure (VP) readings
- Participants
- 147 patients
- Exposure
- Improved alarm algorithm vs dialysis machine alarm for detection of VND
- Outcome
- Alarm activation in VND simulations
Methodological detailsin vitro sham dialysis circuit (Fresenius 2008K), blood substitute
- In vitro sham dialysis circuit (Fresenius 2008K)
- Blood substitute
Main finding
As reported in the abstract
Alarm activation in VND simulations
- VND simulations
- 19
Key notes
- Machine alarm activation time longer than algorithm activation time for all eight machine alarms (range 1-13 seconds)
Observational data: associations, not effects.
Further findings
Upward drift in VP during HD
- VP readings with upward drift
- 16,594 (38%)
- Range of upward drift
- 20-79 mmHg
- Increase in VP, mean±SD
- 11±18 mm Hg
- Within, minutes
- 20±14
Original publication
Detection of Hemodialysis Venous Needle Dislodgment Using Venous Access Pressure Measurements: A Simulation Study. Kidney360. Published 24 March 2023. doi:10. 34067/ kid. 0000000000000093