Sampling from Extracorporeal Circuit: A Step Forward for Dose Monitoring in Continuous Renal Replacement Therapy
Authors' conclusion
The specific value of aspiration flow rate must be proportioned to the prescription of CRRT treatments set by the clinician.
Study context
- Comparison
- Suction rates for sampling from the CRRT extracorporeal circuit
- Outcome
- Optimal suction rate to collect samples from the extracorporeal circuit to measure extracorporeal clearance of a specific solute
Main finding
As reported in the abstract
Optimal suction rate for sampling
- Low suction rate (1 mL/min) ideal for adequate sampling
- Low velocities do not perturb the external circulation and ensure stable circuit pressures
- Higher velocities only with blood flows above 120 mL/min, preferably with appropriate filtration fraction
Original publication
Sampling from Extracorporeal Circuit: A Step Forward for Dose Monitoring in Continuous Renal Replacement Therapy. Blood Purif. Published 22 November 2023. doi:10. 1159/ 000535308
Composed by AI from the published abstract and checked against it. How Briefs are made →