Acute kidney injury

Sampling from Extracorporeal Circuit: A Step Forward for Dose Monitoring in Continuous Renal Replacement Therapy

Blood PurifPublished Mechanistic study

Authors' conclusion

The specific value of aspiration flow rate must be proportioned to the prescription of CRRT treatments set by the clinician.

Study context

Comparison
Suction rates for sampling from the CRRT extracorporeal circuit
Outcome
Optimal suction rate to collect samples from the extracorporeal circuit to measure extracorporeal clearance of a specific solute

Main finding

As reported in the abstract

Optimal suction rate for sampling
  • Low suction rate (1 mL/min) ideal for adequate sampling
  • Low velocities do not perturb the external circulation and ensure stable circuit pressures
  • Higher velocities only with blood flows above 120 mL/min, preferably with appropriate filtration fraction

Original publication

Sampling from Extracorporeal Circuit: A Step Forward for Dose Monitoring in Continuous Renal Replacement Therapy. Blood Purif. Published 22 November 2023. doi:10.1159/000535308

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.