Acute kidney injury

Continuous Renal Replacement Therapy during Extracorporeal Membrane Oxygenation: Circuit Haemodynamics and Circuit Failure

Blood PurifPublished Retrospective cohort study

Authors' conclusion

CRRT circuits used in conjunction with ECMO have a longer circuit life than usual CRRT despite exposure to higher circuit pressures. Markedly elevated access pressures, however, may predict early CRRT circuit failure during ECMO, possibly via progressive membrane thrombosis as evidenced by increased transfilter pressure gradients.

Study context

Population
Continuous renal replacement therapy (CRRT) treatments in two adult ICUs
Exposure
ECMO-CRRT vs non-ECMO-CRRT treatments
Outcome
CRRT circuit life
Methodological detailstwo adult ICUs, 3-year period, 60% training / 40% validation data subsets, time-varying Cox proportional hazard model, classification and regression tree analysis
  • Two adult ICUs
  • 3-year period
  • 60% training / 40% validation data subsets
  • Time-varying Cox proportional hazard model
  • Classification and regression tree analysis

Main finding

As reported in the abstract

CRRT circuit life

P < 0.0001

Observational data: associations, not effects.

Further findings

Predictors of circuit failure (multivariable Cox model)

HR (CI)
First access pressures ≥190 mm HgIndependently associated1.58 (1.09–2.30)
OutcomeHR (CI)
Predictors of circuit failure (multivariable Cox model)Patient weight third tertilevs first tertileIndependently associated1.85 (1.15–2.97)

Circuit pressures

Access, return, prefilter and effluent pressures greater during extracorporeal membrane oxygenation (ECMO)
Higher ECMO flows associated with higher access and return pressures

Predictors of circuit failure (multivariable Cox model)

High access pressures associated with accelerated circuit failure (classification and regression tree analysis)

Transfilter pressure

Access dysfunction associated with stepwise increase in transfilter pressure

Original publication

Continuous Renal Replacement Therapy during Extracorporeal Membrane Oxygenation: Circuit Haemodynamics and Circuit Failure. Blood Purif. Published 19 April 2023. doi:10.1159/000529928

Composed by AI from the published abstract and checked against it. How Briefs are made →