Performance of an interstitial glucose monitoring device in patients with type 1 diabetes during haemodialysis
Authors' conclusion
Use of the FSL2 interstitial glucose monitoring device in HD patients with T1DM is clinically acceptable, even though the accuracy of the device is generally poorer than in studies including non-dialysis patients.
Study context
- Population
- T1DM patients on chronic haemodialysis (HD), during HD sessions565 interstitial fluid glucose (ISFG)/blood glucose (BG) pairs analysed
- Participants
- 12 patients
- Comparison
- FSL2 interstitial glucose vs blood glucose (StatStrip, Accu-Chek Guide) every 15 minutes during 3 HD sessions
- Outcome
- Analytical performance of the FreeStyle Libre 2 (FSL2) sensor: mean absolute relative difference between ISFG and BG
- Follow-up
- During three HD sessions
Methodological detailsprospective, observational, multicentric, Mean age 54 ± 11 years; mean diabetes duration 36.5 ± 11.6 years; dialysis vintage 35 ± 22 months
- Prospective
- Observational
- Multicentric
- Mean age 54 ± 11 years; mean diabetes duration 36.5 ± 11.6 years; dialysis vintage 35 ± 22 months
Performance
As reported in the abstract
Mean absolute relative difference, ISFG vs BG
Key notes
- Interstitial results tend to underestimate blood results
- All values classified as having clinically acceptable error
Further findings
Stability of differences
- Differences remained stable during the dialysis session and were not associated with the ultrafiltration rate
Original publication
Performance of an interstitial glucose monitoring device in patients with type 1 diabetes during haemodialysis. Clin Kidney J. Published 26 February 2024. doi:10. 1093/ ckj/ sfae045