Performance characteristics of a prototype dialysate turbidity monitoring system to detect peritonitis in patients receiving peritoneal dialysis
Authors' conclusion
This proof-of-principle study demonstrates the feasibility and diagnostic accuracy of a prototype dialysate turbidity monitoring system for peritonitis surveillance.
Study context
- Population
- Patients receiving intermittent peritoneal dialysis (PD)983 specimens analysed
- Participants
- 65 patients
- Comparison
- Dialysate turbidity measurements vs International Society of Peritoneal Dialysis (ISPD) diagnostic criteria for peritonitis
- Outcome
- Diagnostic accuracy for peritonitis
Methodological detailssingle-centre, proof-of-principle, prototype bedside dialysate turbidity monitoring system
- Single-centre
- Proof-of-principle
- Prototype bedside dialysate turbidity monitoring system
Performance
As reported in the abstract
Diagnostic accuracy for peritonitis
| Test or model | Sensitivity (unadjusted) | Specificity (unadjusted) |
|---|---|---|
| Turbidity monitoring | 95.2% | 91.5% |
Turbidity monitoring
Sensitivity (unadjusted)95.2%
Specificity (unadjusted)91.5%
Key notes
- 105 samples from patients with peritonitis, 878 without
Further findings
Samples not meeting ISPD criteria
- Most were cases detected before criteria were met or related to active peritonitis treatment and resolution