Kidney transplantation

Patient Self-Testing of Kidney Function at Home, a Prospective Clinical Feasibility Study in Kidney Transplant Recipients

Kidney Int RepPublished Prospective clinical feasibility study

Authors' conclusion

This small feasibility study observed that training selected patients to competently use hand-held devices to self-test kidney function at home is possible. Self-test creatinine results showed good analytical and clinical agreement with standard clinic test results. Self-test potassium results showed poorer agreement with standard clinic test results; however, patient self-use of i-STATs at home was not a statistically significant source of difference between paired potassium test results.

Study context

Population
Stable kidney transplant recipients
Participants
15 recipients
Exposure
Home self-test of capillary blood with i-STAT vs reference clinic test of venous blood (laboratory Siemens Advia Chemistry XPT analyzer)
Outcome
Agreement of blood potassium and creatinine results between home self-tests and clinic reference tests
Methodological detailsobservational, single-center, Bland-Altman and error grid analysis
  • Observational
  • Single-center
  • Bland-Altman and error grid analysis

Main finding

As reported in the abstract

Mean within-patient difference, creatinine

2.25 μmol/l

95% CI −12.13 to 16.81

Between Index self-test and Reference test

Mean within-patient difference, potassium

0.66 mmol/l

95% CI −1.47 to 2.79

Between Index self-test and Reference test

Creatinine pairs judged clinically equivalent
All
Potassium pairs judged clinically equivalent
27 of 40 (67.5%)

Observational data: associations, not effects.

Further findings

Sources of paired test result differences

Biochemical variables associated with potassium measurement in capillary blood suggested as predominant sources of paired differences
  • Planned follow-up analysis

Patient vs nurse i-STAT capillary potassium

Paired patient and nurse i-STAT capillary potassium results not statistically significantly different

Original publication

Patient Self-Testing of Kidney Function at Home, a Prospective Clinical Feasibility Study in Kidney Transplant Recipients. Kidney Int Rep. Published 28 March 2023. doi:10.1016/j.ekir.2023.03.003

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.