Microchip Versus Piezoelectric Point of Care Ultrasonography for Pulmonary and Vena Cava Evaluation in Patients With Acute Kidney Injury
Authors' conclusion
Our study observed a good correlation and agreement between microchip and piezoelectric-based ultrasound modalities in evaluating the presence of pulmonary B-lines and inferior vena cava (IVC) dynamics in patients with AKI. Microchip ultrasound, in addition to being portable, user-friendly, and cost-effective, consistently delivers bedside parameters for volume assessment that are comparable to those obtained through POCUS using piezoelectric transducer-based ultrasonography.
Study context
- Population
- Critically ill intensive care unit (ICU) patients with acute kidney injury (AKI) receiving kidney replacement therapy (KRT)
- Participants
- 50 patients
- Comparison
- Microchip vs piezoelectric point-of-care ultrasonography (POCUS)
- Outcome
- Correlation and agreement between the 2 scanner modalities for lung B-lines (8 quadrants) and inferior vena cava (IVC) collapsibility index (IVCi)
- Follow-up
- Before KRT initiation (T0) and 60 minutes after start (T60)
Methodological detailsrandomized order of scanners, sequential examinations at 2 time points
- Randomized order of scanners
- Sequential examinations at 2 time points
Main finding
As reported in the abstract
T0 and T60
- B-lines correlation at T0
- ⍴ = 0.96 (P < 0.001)
- B-lines correlation at T60
- ⍴ = 0.93 (P < 0.001)
- IVCi correlation at T0
- ρ = 0.70 (P < 0.001)
- IVCi correlation at T60
- ρ = 0.87 (P < 0.001)
- IVCi Bland-Altman bias at T0
- +2.69 (SD: 10.6) (95% CI: −18.13 to +23.52)
- IVCi Bland-Altman bias at T60
- 3.28 (SD: 10.23) (95% CI: −16.77 to +23.34)
- B-lines Bland-Altman bias at T0
- −0.3 (SD: 2.73), 95% CI −5.66 to +5.06
- B-lines Bland-Altman bias at T60
- 0.2 (SD: 3.23), 95% CI −6.14 to +6.54
Key notes
- Bland-Altman plots showed agreement between methods for IVCi and B-lines
- Intervals as printed with the Bland-Altman bias
Original publication
Microchip Versus Piezoelectric Point of Care Ultrasonography for Pulmonary and Vena Cava Evaluation in Patients With Acute Kidney Injury. Kidney Int Rep. Published 26 November 2023. doi:10. 1016/ j. ekir. 2023. 11. 019
Composed by AI from the published abstract and checked against it. How Briefs are made →