Intravascular Ultrasound Is More Accurate Than Angiography in Arteriovenous Vascular Access Lesions
Authors' conclusion
Geometric complexity directly contributes to modality-specific error. Angiography systematically underestimates lumen dimensions in complex, stenotic regions, while IVUS detects them with higher fidelity and preserves accuracy. These findings establish IVUS as the more reliable modality for evaluating AV access dysfunction and support its integration into routine practice in guiding intervention for AV access stenosis.
Study context
- Population
- Clinically relevant arteriovenous (AV) access geometries modeled with computer-aided design (CAD) and fabricated by 3D printing
- Comparison
- Intravascular ultrasound (IVUS) vs contrast angiography, against CAD dimensions
- Outcome
- Lumen diameter measurement error compared with CAD dimensions
Methodological detailsthree-dimensional-printed vascular conduits as reference standards, finite element-based Gaussian curvature mapping, linear mixed-effect modeling
- Three-dimensional-printed vascular conduits as reference standards
- Finite element-based Gaussian curvature mapping
- Linear mixed-effect modeling
Performance
As reported in the abstract
Measurement error vs CAD
- Angiography consistently underestimated lumen diameter in regions of high Gaussian curvature, error increasing in curved or tortuous lesions
- IVUS closely approximated CAD measurements, retaining accuracy even in severe stenoses
- IVUS vs Angiography: P < 0.05
Further findings
Mild stenoses (<50%) and aneurysmal dilatations
- Both modalities performed comparably
Original publication
Intravascular Ultrasound Is More Accurate Than Angiography in Arteriovenous Vascular Access Lesions. Kidney360. Published 13 February 2026. doi:10. 34067/ kid. 0000001128
Composed by AI from the published abstract and checked against it. How Briefs are made →