Intradialytic hypotension is severely underreported by intermittent blood pressure readings
Authors' conclusion
First, the occurrence of IDH is severely underestimated with intermittent BP monitoring. Second, FC-BP appears unsuitable for routine intradialytic BP monitoring due to its poor accuracy, precision, and concordance.
Study context
- Population
- Chronic dialysis patients, BP measured during four sessions
- Participants
- 40 patients
- Comparison
- Intermittent brachial artery cuff BP (every 15 minutes) vs continuous finger cuff BP monitoring
- Outcome
- Intradialytic hypotension (IDH) detection (SBP <90 or <100 mmHg, depending on pre-dialysis SBP)
Methodological detailsdevice validation, Bland-Altman analyses, four-quadrant plots
- Device validation
- Bland-Altman analyses
- Four-quadrant plots
Performance
As reported in the abstract
IDH episodes detected · 106 hours
| Test or model | Value | Readings | Measurements |
|---|---|---|---|
| Brachial artery cuff-blood pressureBAC-BP | 20 events | 144 | — not stated in the abstract |
| FC-BP | 441 events | — not stated in the abstract | 18 163 |
Value
Brachial artery cuff-blood pressureBAC-BP20 events
FC-BP441 events
Readings
Brachial artery cuff-blood pressureBAC-BP144
FC-BP— not stated in the abstract
Measurements
Brachial artery cuff-blood pressureBAC-BP— not stated in the abstract
FC-BP18 163
Further findings
FC-BP validation vs BAC-BP (1221 paired readings)
- Systolic BP (SBP) mean bias ± limits of agreement
- −13.3 ± 44.5 mmHg
- Diastolic BP mean bias ± limits of agreement
- −13.3 ± 25.5 mmHg
- Paired FC-BP readings meeting acceptable accuracy criterion
- 27.7%
- Concordance rates (SBP and diastolic BP, respectively)
- 34.3% and 36.3%
- Acceptable accuracy: mean SBP bias ≤10 mmHg
Original publication
Intradialytic hypotension is severely underreported by intermittent blood pressure readings. Clin Kidney J. Published 11 November 2025. doi:10. 1093/ ckj/ sfaf337