A novel system to continuously estimate intradialytic blood pressure in real time
Authors' conclusion
A blood pressure estimator that runs in real time during haemodialysis using pressure wave sensors in the extracorporeal circuit and avoiding additional sensor-burden on patients has good performance in tracking intradialytic SBP when compared against brachial cuff measurements, supporting its further development and larger scale testing.
Study context
- Population
- HD patients with a functioning AVF, studied throughout two 4-h HD sessions522 systolic blood pressure (SBP) comparison data points
- Participants
- 21 patients
- Comparison
- Real-time SBP estimator vs brachial cuff SBP
- Outcome
- SBP
Methodological detailssingle-centre, observational
- Single-centre
- Observational
Performance
As reported in the abstract
Absolute mean difference
−1.9 ± 16 mmHg
low; no evidence of systematic bias between measurements
R
0.825
P < 0.001
significantly associated
- Estimator values within 30% of the matched brachial cuff value
- 95%
- Estimator values within 10% of the cuff value
- 66%
Key notes
- Across 522 SBP comparison data points
Original publication
A novel system to continuously estimate intradialytic blood pressure in real time. Nephrol Dial Transplant. Published 24 March 2025. doi:10. 1093/ ndt/ gfaf058