Hemodialysis

A novel system to continuously estimate intradialytic blood pressure in real time

Nephrol Dial TransplantPublished Diagnostic accuracy study

Authors' conclusion

A blood pressure estimator that runs in real time during haemodialysis using pressure wave sensors in the extracorporeal circuit and avoiding additional sensor-burden on patients has good performance in tracking intradialytic SBP when compared against brachial cuff measurements, supporting its further development and larger scale testing.

Study context

Population
HD patients with a functioning AVF, studied throughout two 4-h HD sessions522 systolic blood pressure (SBP) comparison data points
Participants
21 patients
Comparison
Real-time SBP estimator vs brachial cuff SBP
Outcome
SBP
Methodological detailssingle-centre, observational
  • Single-centre
  • Observational

Performance

As reported in the abstract

Absolute mean difference

−1.9 ± 16 mmHg

low; no evidence of systematic bias between measurements

R

0.825

P < 0.001

significantly associated

Estimator values within 30% of the matched brachial cuff value
95%
Estimator values within 10% of the cuff value
66%