Increased Longevity of a Novel Gas Exchanger System for Low-Flow Veno-Venous Extracorporeal CO2 Removal in Acute Hypercapnic Respiratory Failure

Blood PurifPublished Retrospective cohort study

Authors' conclusion

Using a high-longevity gas exchanger system, membrane lifetime in low-flow ECCO2R could be extended in comparison to previous systems but remained below 120 h, especially in spontaneously breathing patients. Extracorporeal VCO2 remained constant throughout gas exchanger system runtime and was consistent with removal of approximately 50% of expected CO2 production, enabling lung protective ventilation despite hypercapnic respiratory failure.

Study context

Population
Patients with pH ≤ 7.25 and/or PaCO2 ≥9 kPa exceeding lung protective ventilation limits, or with respiratory exhaustion during spontaneous breathing, in acute hypercapnic respiratory failure
Participants
70 patients
Exposure
High-longevity ProLUNG gas exchanger system vs original gas exchanger
Outcome
Gas exchanger runtime (hypothesis: runtimes up to 120 h)
Methodological detailsretrospective analysis, low-flow veno-venous extracorporeal CO2 removal (ECCO2R), 9,806 treatment-hours recorded
  • Retrospective analysis
  • Low-flow veno-venous ECCO2R
  • 9,806 treatment-hours recorded

Main finding

As reported in the abstract

Gas exchanger runtime
High-longevity system increased gas exchanger runtime in ventilated patients
from 29 ± 16 to 48 ± 36 h
High-longevity system increased gas exchanger runtime in awake patients
from 22 ± 14 to 31 ± 31 h
Increase in runtime with the high-longevity system
p < 0.0001
Longer runtime in ventilated than in awake patients
p < 0.01

Observational data: associations, not effects.

Further findings

Sweep-gas VCO2

VCO2 remained constant
86 ± 34 mL/min (p = 0.11)

PaCO2, overall · within 1 h

PaCO2 decreased
from 9.1 ± 2.0 to 7.9 ± 1.9 kPa (p < 0.001)

Tidal volume · up to 120 h

Tidal volume maintained
5.4 ± 1.8 versus 5.7 ± 2.2 mL/kg (p = 0.60)

Peak airway pressure

Peak airway pressure reduced
from 31.1 ± 5.1 to 27.5 ± 6.8 mbar (p < 0.01)

Original publication

Increased Longevity of a Novel Gas Exchanger System for Low-Flow Veno-Venous Extracorporeal CO2 Removal in Acute Hypercapnic Respiratory Failure. Blood Purif. Published 2 November 2022. doi:10.1159/000526582

Composed by AI from the published abstract and checked against it. How Briefs are made →