Increased Longevity of a Novel Gas Exchanger System for Low-Flow Veno-Venous Extracorporeal CO2 Removal in Acute Hypercapnic Respiratory Failure
Authors' conclusion
Using a high-longevity gas exchanger system, membrane lifetime in low-flow ECCO2R could be extended in comparison to previous systems but remained below 120 h, especially in spontaneously breathing patients. Extracorporeal VCO2 remained constant throughout gas exchanger system runtime and was consistent with removal of approximately 50% of expected CO2 production, enabling lung protective ventilation despite hypercapnic respiratory failure.
Study context
- Population
- Patients with pH ≤ 7.25 and/or PaCO2 ≥9 kPa exceeding lung protective ventilation limits, or with respiratory exhaustion during spontaneous breathing, in acute hypercapnic respiratory failure
- Participants
- 70 patients
- Exposure
- High-longevity ProLUNG gas exchanger system vs original gas exchanger
- Outcome
- Gas exchanger runtime (hypothesis: runtimes up to 120 h)
Methodological detailsretrospective analysis, low-flow veno-venous extracorporeal CO2 removal (ECCO2R), 9,806 treatment-hours recorded
- Retrospective analysis
- Low-flow veno-venous ECCO2R
- 9,806 treatment-hours recorded
Main finding
As reported in the abstract
Gas exchanger runtime
- High-longevity system increased gas exchanger runtime in ventilated patients
- from 29 ± 16 to 48 ± 36 h
- High-longevity system increased gas exchanger runtime in awake patients
- from 22 ± 14 to 31 ± 31 h
- Increase in runtime with the high-longevity system
- p < 0.0001
- Longer runtime in ventilated than in awake patients
- p < 0.01
Key notes
- Retrospective analysis
Observational data: associations, not effects.
Further findings
Sweep-gas VCO2
- VCO2 remained constant
- 86 ± 34 mL/min (p = 0.11)
PaCO2, overall · within 1 h
- PaCO2 decreased
- from 9.1 ± 2.0 to 7.9 ± 1.9 kPa (p < 0.001)
Tidal volume · up to 120 h
- Tidal volume maintained
- 5.4 ± 1.8 versus 5.7 ± 2.2 mL/kg (p = 0.60)
Peak airway pressure
- Peak airway pressure reduced
- from 31.1 ± 5.1 to 27.5 ± 6.8 mbar (p < 0.01)