Human factors testing in a novel automated peritoneal dialysis system
Authors' conclusion
In this summative human factors usability study, the Archimedes APD system usability was found to be safe, accessible and easy for its intended users and use environments after the relatively short training period compared to the status quo.
Study context
- Population
- Current or former peritoneal dialysis (PD) patients and dialysis nurses using a novel gravity-based automated peritoneal dialysis (APD) device (Archimedes™)
- Participants
- 15 vs 15
- Comparison
- Patients vs Nurses
- Patientsn = 15
- Nursesn = 15
- Outcome
- Success of critical Use Scenario tasks (Simulated Use and Knowledge tasks)
- Follow-up
- After training decay period
Methodological detailstraining decay period, 2 h training, then training decay period, then evaluation
- Training decay period
- 2 h training, then training decay period, then evaluation
Main finding
As reported in the abstract
Simulated Use task success
- Simulated Use tasks successful across all users
- 97.3%
Further findings
Knowledge task success
Original publication
Human factors testing in a novel automated peritoneal dialysis system. Perit Dial Int. Published 14 August 2026. doi:10. 1177/ 08968608261476545