Blood PurifPublished Prospective crossover study

  • 2 patients terminated prematurely before study intervention, 1 due to a protocol violation

Authors' conclusion

The SILENCIA cycler delivers adequate urea clearance and UF. Importantly, sleep quality improved, possibly related to less caution messages and alarms.

Study context

Population
Patients on automated peritoneal dialysis (APD)13 evaluated for Kt/V and ultrafiltration (UF)
Participants
16 patients
Exposure
SILENCIA cycler vs current cycler for APD
Outcome
Total Kt/Vurea and UF volume
Methodological detailscross-over, two 2-week study periods, 3-week training phase
  • Cross-over
  • Two 2-week study periods
  • 3-week training phase

Main finding

As reported in the abstract

  • Neither Kt/Vurea nor UF differed significantly between control and SILENCIA cyclers
  • 13 patients evaluated

Observational data: associations, not effects.

Further findings

Average reported sleep time · spread printed as ±; kind assumed SD — SILENCIA
average 8.0 ± 1.6 h
  • Average reported sleep time: PD-NIGHT average 5.9 ± 1.8 h
  • Average reported sleep time: HomeChoice Pro average 7.2 ± 2.1 h

Sleep quality · after the 2-week phase with SILENCIA

Sleep quality improved
5 of 10 patients
Sleep quality unchanged vs previous cycler
other patients
  • 10 patients answered the questionnaire

Satisfaction with new cycler

All patients much or very much satisfied with the new cycler

Original publication

A New Cycler for Automated Peritoneal Dialysis to Provide Efficient Dialysis and Improved Sleep Quality. Blood Purif. Published 16 June 2023. doi:10.1159/000530464

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.