Handheld Dynamometry Testing During Dialysis: Intrarater and Interrater Reliability Study
Authors' conclusion
The assessment of the hip flexion and abduction strength during dialysis with a handheld dynamometer is reliable and safe. Normalized data provides more reliable results.
Study context
- Population
- Subjects undertaking HD for at least 3 monthsmedian age 67 years
- Participants
- 56 subjects
- Exposure
- Raters A vs Raters B
- Outcome
- Intrarater and interrater reliability of handheld dynamometry testing of lower limb (hip and ankle) muscle strength during HD
Methodological detailsintrarater and interrater reliability study, Bland-Altman plots, intraclass correlation coefficient, minimal detectable change
- Intrarater and interrater reliability study
- Bland-Altman plots
- Intraclass correlation coefficient
- Minimal detectable change
Main finding
As reported in the abstract
Relative reliability, high to very high (intraclass correlation coefficients)
0.75-0.98
Mean difference between trials, raters A
1.0-19.3 N
Mean difference between trials, raters B
1.3 - 11.4 N
Mean difference between raters
2.4 to 13.5 N
Key notes
- Intrarater minimal detectable change increased for ankle muscles; interrater values were wider
- Bland-Altman plot: systematic bias for raters A in ankle plantar flexion
Observational data: associations, not effects.