Hemodialysis

Efficacy and Safety of the Medium Cut-Off Elisio HX Dialyzer

Blood PurifPublished Prospective crossover study

Authors' conclusion

The new MCO dialyzer, Elisio HX, performs with excellent behavior and tolerance. It represents an upgrade compared to their predecessor and is very close to the removal capacity of HDF treatment.

Study context

Population
Patients on dialysis5 sessions each
Participants
18 patients
Exposure
Elisio HX19 in HDx vs FX80 Cordiax in Hemodialysis (HD), Elisio H19 in HD, Theranova 400 in HDx and FX80 Cordiax in Hemodiafiltration (HDF)
Outcome
Reduction ratios of urea, creatinine, ß2-microglobulin, myoglobin, kappa FLC, prolactin, α1-microglobulin, α1-acid glycoprotein, lambda FLC and albumin
Methodological details5 dialysis sessions per patient
  • 5 dialysis sessions per patient

Main finding

As reported in the abstract

Reduction ratios of small, medium and large molecules
  • No difference between modalities for small molecules
  • HDx and HDF significantly more efficient than HD for medium and large molecule removal
  • Elisio HX19 in HDx similar to Theranova 400, superior to both dialyzers in HD, slightly lower than HDF

Observational data: associations, not effects.

Further findings

Global removal score (GRS)

GRS significantly higher with HDx than with HD
Highest GRS with the helixone dialyzer in HDF

Safety

Dialysate albumin loss

HD dialyzers
less than 1 g
HDx and HDF
between 1.5 and 2.5 g

Original publication

Efficacy and Safety of the Medium Cut-Off Elisio HX Dialyzer. Blood Purif. Published 12 May 2022. doi:10.1159/000524644