Efficacy and Safety of the Medium Cut-Off Elisio HX Dialyzer
Authors' conclusion
The new MCO dialyzer, Elisio HX, performs with excellent behavior and tolerance. It represents an upgrade compared to their predecessor and is very close to the removal capacity of HDF treatment.
Study context
- Population
- Patients on dialysis5 sessions each
- Participants
- 18 patients
- Exposure
- Elisio HX19 in HDx vs FX80 Cordiax in Hemodialysis (HD), Elisio H19 in HD, Theranova 400 in HDx and FX80 Cordiax in Hemodiafiltration (HDF)
- Outcome
- Reduction ratios of urea, creatinine, ß2-microglobulin, myoglobin, kappa FLC, prolactin, α1-microglobulin, α1-acid glycoprotein, lambda FLC and albumin
Methodological details5 dialysis sessions per patient
- 5 dialysis sessions per patient
Main finding
As reported in the abstract
Reduction ratios of small, medium and large molecules
- No difference between modalities for small molecules
- HDx and HDF significantly more efficient than HD for medium and large molecule removal
- Elisio HX19 in HDx similar to Theranova 400, superior to both dialyzers in HD, slightly lower than HDF
Observational data: associations, not effects.
Further findings
Global removal score (GRS)
- GRS significantly higher with HDx than with HD
- Highest GRS with the helixone dialyzer in HDF
Safety
Dialysate albumin loss
- HD dialyzers
- less than 1 g
- HDx and HDF
- between 1.5 and 2.5 g
Original publication
Efficacy and Safety of the Medium Cut-Off Elisio HX Dialyzer. Blood Purif. Published 12 May 2022. doi:10. 1159/ 000524644