Detection and prognostic relevance of phospholipase A2 receptor (PLA2R) epitopes in idiopathic membranous nephropathy: a simultaneous quantitative multiplex suspension array detection method
Authors' conclusion
A multiplex suspension assay was developed for the simultaneous quantitative detection of anti-CysR/CTLD1/CTLD678-IgG4 antibodies. The epitope migration sequence of PLA2R molecules during disease progression may not follow a simple linear rule. Among them, the epitope CTLD1 is likely to exert the most significant influence on patient remission.
Study context
- Population
- Patients with idiopathic MN; healthy individuals
- Outcome
- Assay performance and epitope findings
Methodological detailsmultiplex suspension array
- Multiplex suspension array
Performance
As reported in the abstract
- All three epitopes detected simultaneously within
- 2 h
- Intra-assay precision
- 4.74% to 9.48%
- Inter-assay precision
- 5.13% and 13.92%
- Recovery rates
- 90% and 100%
- Cut-off values CysR, CTLD1, CTLD678 (healthy vs patients)
- 6.30, 12.38 and 10.06 Ru/mL
- Positive rates CysR, CTLD1, CTLD678
- 100%, 34% and 75%
- Group 3 (CTLD1 response) share of the 12 patients not in remission
- 73%
- CTLD1 concentration above which prognosis may be poorer
- 42.76 Ru/mL
Key notes
- Human IgA antibody did not cause a cross-reaction
Original publication
Detection and prognostic relevance of PLA2R epitopes in idiopathic membranous nephropathy: a simultaneous quantitative multiplex suspension array detection method. Clin Kidney J. Published 13 January 2025. doi:10. 1093/ ckj/ sfaf010
Composed by AI from the published abstract and checked against it. How Briefs are made →