Clinical performance and microstructural assessment of hemodialysis catheters for children on maintenance dialysis
Authors' conclusion
Pediatric CVCs are associated with a high rate of dysfunction, leading to replacement in 44% of patients. Interestingly, diameters ≥ 10Fr (12.5Fr, ≥ 14Fr) were not associated with improved survival, whereas 8Fr catheters showed a non-significant trend toward increased occlusion and poorer catheter survival, suggesting a possible threshold effect. Micro-CT and histology confirmed the presence, site and extent of thrombi. These data can inform development and modeling of new CVC designs.
Study context
- Population
- Children aged 0–18 years receiving maintenance hemodialysis (HD) via CVC288 catheters
- Outcome
- CVC-related outcomes: complication rates and access survival
Methodological detailsthree pediatric dialysis centers, histological sectioning
- Three pediatric dialysis centers
- Histological sectioning
Main finding
As reported in the abstract
- CVC dysfunction
- 54.2%
- Dysfunction due to infection
- 40.6%
- Dysfunction due to line occlusion
- 24.5%
- Median CVC survival
- 215.5 days
- CVC replacements in 72 children
- 133
- Infection rates per 1000 catheter-days, depending on CVC size
- 0.20 to 1.82
Key notes
- Larger CVCs (≥ 14Fr) did not show better access survival vs 10-Fr models
- May be influenced by design differences
Further findings
Micro-CT and histology of explanted central venous catheters (CVCs)
- Explanted CVCs examined by micro-CT
- 16
- Occlusions found by micro-CT
- all
- Histologically confirmed thrombi
- 7
- Thrombi rich in fibrin, predominantly at side holes and catheter tip
Original publication
Clinical performance and microstructural assessment of hemodialysis catheters for children on maintenance dialysis. Pediatr Nephrol. Published 13 August 2026. doi:10. 1007/ s00467-026-07500-6
Composed by AI from the published abstract and checked against it. How Briefs are made →