Electrolytes, acid–base & stones

Clinical Evaluation of a Disposable Test Strip for Potassium Level Measurement in Patients Treated with Maintenance Hemodialysis

Clin J Am Soc NephrolPublished Diagnostic accuracy study

Authors' conclusion

POCT for potassium level using the disposable test strip could be used for self-monitoring of blood potassium levels in clinical practice.

Study context

Population
Patients undergoing maintenance HD
Participants
40 patients
Comparison
Disposable potassium test strip vs central laboratory equipment
Outcome
Agreement between potassium levels measured by disposable test strips and by central laboratory equipment
Methodological detailsexperimental testing with standard potassium solutions, Passing-Bablok regression, Bland-Altman analysis
  • Experimental testing with standard potassium solutions
  • Passing-Bablok regression
  • Bland-Altman analysis

Performance

As reported in the abstract

Agreement of capillary-blood test strip potassium with central laboratory serum potassium

Passing-Bablok regression slope

1.04

95% CI 0.97 to 1.12

Further findings

Experimental performance with potassium solutions

Coefficient of variation across potassium concentrations 2 to 9 mM
<5%
High degree of agreement with inductively coupled plasma atomic emission spectroscopy

Original publication

Clinical Evaluation of a Disposable Test Strip for Potassium Level Measurement in Patients Treated with Maintenance Hemodialysis. Clin J Am Soc Nephrol. Published 5 December 2025. doi:10.2215/cjn.0000000934

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.