Use of the B. Braun Hemodialysis Machine to Provide Home Hemodialysis
Authors' conclusion
B. Braun HD machine is another HD device that can be considered to provide home HD. Our study has shown that B. Braun machine had higher spKt/V in addition to standard weekly Kt/V and URR when compared to the NxStage machine. Larger, well-designed, and well-powered studies are needed to confirm these results.
Study context
- Population
- Patients receiving home hemodialysis (HD)
- Participants
- 27 vs 36
- Exposure
- Home HD with B. Braun Dialog+ HD machine vs NxStage System One machine
- B. Braunn = 27
- NxStagen = 36reference
- Outcome
- Feasibility of using the B. Braun Dialog+ HD machine to provide home HD
Main finding
As reported in the abstract
Single treatment spKt/V
B. Braun
1.48
± 0.27
NxStage
1.14
± 0.32
P < 0.001
Key notes
- Better with B. Braun
Observational data: associations, not effects.
Further findings
| Outcome | B. Braun | NxStage | P value |
|---|---|---|---|
| Standard weekly Kt/VBetter with B. Braun | 4.44 ± 0.82 | 2.65 ± 0.57 | P < 0.001 |
| URRBetter with B. Braun | 71.4 ± 6.2 | 59.2 ± 9.1 | P < 0.001 |
| HD treatment durationSignificantly shorter with B. Braun | 3.48 ± 0.32 | 4.58 ± 1.52 | P < 0.001 |
| Ultrafiltration rateSignificantly larger with B. Braun | 1.83 ± 0.64 | 1.18 ± 0.66 | P < 0.001 |
| Bicarbonate levelLower with B. Braun | 23.5 ± 2.0 | 24.9 ± 2.8 | P = 0.021 |
Other laboratory data
- No significant differences in average albumin, BUN, calcium, creatinine, hemoglobin, phosphorus, platelets, potassium, WBC, or PTH
Original publication
Use of the B. Braun Hemodialysis Machine to Provide Home Hemodialysis. Hemodial Int. Published 25 May 2025. doi:10. 1111/ hdi. 13254
Composed by AI from the published abstract and checked against it. How Briefs are made →