Unveiling a new era with liquid chromatography coupled with mass spectrometry to enhance parathyroid hormone measurement in patients with chronic kidney disease
Authors' conclusion
The recalibration successfully reduced inter-assay variability with harmonization of PTH measurements across different assays. By recalibrating the assays based on the WHO 95/646 International Standard, we demonstrated the feasibility for standardizing PTH measurement results and adopting common reference ranges for PTH assays, facilitating a more consistent interpretation of PTH values.
Study context
- Population
- Pooled plasma samples for parathyroid hormone (PTH) measurement in CKD5 PTH immunoassays
- Comparison
- PTH immunoassays recalibrated with LC-MS/MS as reference
- Outcome
- Impact of recalibrating PTH immunoassays by an LC-MS/MS reference method
Methodological detailsrecalibration of five PTH immunoassays against LC-MS/MS, World Health Organization (WHO) 95/646 International Standard, pooled plasma samples
- Recalibration of five PTH immunoassays against LC-MS/MS
- WHO 95/646 International Standard
- Pooled plasma samples
Main finding
As reported in the abstract
Inter-assay variability of PTH
- Recalibration reduced inter-assay variability, harmonising PTH measurements across assays
Further findings
- Robustness of standardisation evaluated over time with different sets of samples
Original publication
Unveiling a new era with liquid chromatography coupled with mass spectrometry to enhance parathyroid hormone measurement in patients with chronic kidney disease. Kidney Int. Published 31 October 2023. doi:10. 1016/ j. kint. 2023. 09. 033