Ultrasound localization microscopy of the human kidney allograft on a clinical ultrasound scanner
Authors' conclusion
Thus, this proof-of-concept study has shown ULM to be a promising imaging technique for qualitative and quantitative microvascular assessment.
Study context
- Population
- Human kidney allograftsmethods optimized on 35 transplants
- Participants
- 7 participants
- Comparison
- Ultrasound Localization Microscopy (ULM) vs Color Doppler, Advanced Dynamic Flow and Superb Microvascular Imaging
- Outcome
- Microvascular visualisation by ULM
Methodological detailsproof-of-concept, standard low frame rate clinical ultrasound scanner, contrast agent, Acquisition parameters derived from CEUS examinations by increasing the duration of the recorded clip at the same plane
- Proof-of-concept
- Standard low frame rate clinical ultrasound scanner
- Contrast agent
- Acquisition parameters derived from CEUS examinations by increasing the duration of the recorded clip at the same plane
Main finding
As reported in the abstract
Microvascular visualisation by ULM
- ULM identified vessels thinner vs Doppler modes
- two to four times
- Mean ULM smallest analyzable vessel cross section
- 0.3 ± 0.2 mm
Key notes
- Interlobar, arcuate, cortical radial vessels and part of the medullary organization visible on ULM density maps
- Medullary vasa recta seen but not as clear as cortical vessels
More notes
- ULM provided quantitative information on blood velocities in the cortical area
- Limitations due to movement and saturation artifacts