Hemodialysis

The use of lung ultrasound in evaluation of extravascular lung water in hemodialysis patients: Systematic review and meta‐analysis

Hemodial IntPublished Systematic review and meta-analysis

Authors' conclusion

These findings indicate the effectiveness of lung ultrasound in detection of volume overload and assessment of response to ultrafiltration in hemodialysis patients.

Study context

Population
HD patients assessed with lung US
Participants
2,301 patients33 studies
Comparison
Post-HD vs pre-HD
Outcome
Change in number of B-lines post-HD vs pre-HD
Methodological detailssearches of PubMed, WOS and Scopus
  • Searches of PubMed, WOS and Scopus

Pooled result

Number of B-lines

Mean difference

8.30

95% CI 3.55–13.05

As reported in the abstract

Further pooled results

OutcomeMean difference (95% CI)
Inspiratory IVC diameterReduction post-HD2.32 (0.31–4.33)
Expiratory IVC diameterReduction post-HD4.05 (2.44–5.65)

Correlation of B-lines with maximum IVC diameter

Correlation coefficient, pre-HD
0.39
Correlation coefficient, post-HD
0.32
  • Significant positive correlation

Original publication

The use of lung ultrasound in evaluation of extravascular lung water in hemodialysis patients: Systematic review and meta‐analysis. Hemodial Int. Published 27 February 2024. doi:10.1111/hdi.13141