The Clinical Impact of Urinalysis Screened by Automated Microscopy Compared to Reference Manual Analysis

Am J NephrolPublished Diagnostic accuracy study

Authors' conclusion

Employing automated urinalysis seems to be accurate for detecting hematuria and leukocyturia, as well as for screening patients without kidney diseases. However, clinical laboratories attending complex patients should employ personalized strategies to help decide when to perform manual review, thus avoiding misleading urinalysis results.

Study context

Population
Urine samples randomly selected from routine laboratory specimens at a university hospital, patients with and without kidney disease503 samples
Comparison
Laboratory automated urinalysis vs reference manual microscopy
Outcome
Accuracy of laboratory urinary sediment analysis vs standardized microscopic review
Methodological detailssingle university hospital, random sampling of routine specimens, double-blinded reference analysis
  • Single university hospital
  • Random sampling of routine specimens
  • Double-blinded reference analysis

Performance

As reported in the abstract

Accuracy of laboratory urinalysis vs reference analysis
Test or modelOverall agreementPathological casts, lipiduria, renal tubular epithelial cells: sensitivityPathological casts, lipiduria, renal tubular epithelial cells: specificityHyaline casts: sensitivityHyaline casts: specificityDysmorphic RBCs: sensitivityDysmorphic RBCs: specificityHematuria: sensitivityHematuria: specificityHematuria: ICCHematuria: RLeukocyturia: sensitivityLeukocyturia: specificityLeukocyturia: ICCLeukocyturia: R
Laboratory42.1%<50%>98%50.4%80.9%62.3%96.2%86.1%82.3%0.7030.82884.9%95.1%0.8070.861

Laboratory

Overall agreement42.1%
Pathological casts, lipiduria, renal tubular epithelial cells: sensitivity<50%
Pathological casts, lipiduria, renal tubular epithelial cells: specificity>98%
Hyaline casts: sensitivity50.4%
Hyaline casts: specificity80.9%
Dysmorphic RBCs: sensitivity62.3%
Dysmorphic RBCs: specificity96.2%
Hematuria: sensitivity86.1%
Hematuria: specificity82.3%
Hematuria: ICC0.703
Hematuria: R0.828
Leukocyturia: sensitivity84.9%
Leukocyturia: specificity95.1%
Leukocyturia: ICC0.807
Leukocyturia: R0.861

Further findings

OutcomeP value
Identification of elements suggestive of nephropathyLaboratory, manually reviewed samples vs ReferenceP = 0.503
  • Identification of elements suggestive of nephropathy: only manually reviewed samples statistically similar to reference

Accuracy in patients with kidney disease and in samples manually reviewed by the laboratory

Accuracy for each urinary element comparable to the overall sample findings
  • Kidney disease N = 248; manually reviewed N = 115

Original publication

The Clinical Impact of Urinalysis Screened by Automated Microscopy Compared to Reference Manual Analysis. Am J Nephrol. Published 16 October 2024. doi:10.1159/000541561

Composed by AI from the published abstract and checked against it. How Briefs are made →