Successful Application of a Hemoperfusion Machine in Separated System Continuous Renal Replacement Therapy (CRRT): Evaluating Effectiveness and Safety in Critically Ill Patients With Acute Kidney Injury (AKI) in Resource‐Limited Settings
Authors' conclusion
Separated system CRRT using a hemoperfusion machine provides a simplified system to operate and is proven to be effective and safe in real-life practice, especially in resource-limited areas.
Study context
- Population
- Critically ill patients with AKI, January 2015 to December 2021
- Participants
- 129 patients
- Outcome
- Effectiveness and safety of separated CRRT applied from a hemoperfusion machine
Methodological detailssingle-group, multivariate logistic regression analysis
- Single-group
- Multivariate logistic regression analysis
Main finding
As reported in the abstract
In-hospital mortality rate
68.9%
Serious complications including circuit explosion and air embolism
0
Mean prescribed CRRT dose (mL/kg/h)
22.4 ± 3.1
Key notes
- All patients received pre- and postdilution continuous veno-venous hemofiltration (CVVH)
Further findings
Baseline at CRRT initiation
- SOFA score
- 12.6 ± 3.8
- Fluid accumulation
- 3900 mL (622-8172 mL)
Safety
- No serious complications including circuit explosion and air embolism