Seraph 100 Microbind Affinity Blood Filter Does Not Clear Antibiotics: An Analysis of Antibiotic Concentration Data from PURIFY-OBS
Authors' conclusion
The Seraph 100 did not significantly clear any measured antibiotic in clinical samples. These data give further evidence to suggest that these therapies may be safely administered to critically ill patients and will not impact concentrations of administered antibiotics.
Study context
- Population
- Patients undergoing treatment with the Seraph 100 Microbind Affinity Blood Filter and receiving antibiotics concurrently17 analysed
- Participants
- 34 patients
- Outcome
- Antibiotic clearance due to the Seraph 100 device (CLSeraph, L/h), from pre- and post-filter samples at 1 h and 4 h
- Follow-up
- 1 h and 4 h after treatment initiation
Methodological detailsmulticentre, 6 sites in the USA
- Multicentre
- 6 sites in the USA
Main finding
As reported in the abstract
Antibiotic clearance due to the Seraph 100 device (CLSeraph)
Mean CLSeraph for the antibiotics investigated ranged from
−0.57 to 0.47 L/h
Key notes
- No antibiotic had a CLSeraph statistically significant from 0
- 7 antibiotics and one beta-lactamase inhibitor
Original publication
Seraph 100 Microbind Affinity Blood Filter Does Not Clear Antibiotics: An Analysis of Antibiotic Concentration Data from PURIFY-OBS. Blood Purif. Published 13 January 2024. doi:10. 1159/ 000531951
Composed by AI from the published abstract and checked against it. How Briefs are made →