Seraph 100 Microbind Affinity Blood Filter Does Not Clear Antibiotics: An Analysis of Antibiotic Concentration Data from PURIFY-OBS

Blood PurifPublished Prospective observational study

Authors' conclusion

The Seraph 100 did not significantly clear any measured antibiotic in clinical samples. These data give further evidence to suggest that these therapies may be safely administered to critically ill patients and will not impact concentrations of administered antibiotics.

Study context

Population
Patients undergoing treatment with the Seraph 100 Microbind Affinity Blood Filter and receiving antibiotics concurrently17 analysed
Participants
34 patients
Outcome
Antibiotic clearance due to the Seraph 100 device (CLSeraph, L/h), from pre- and post-filter samples at 1 h and 4 h
Follow-up
1 h and 4 h after treatment initiation
Methodological detailsmulticentre, 6 sites in the USA
  • Multicentre
  • 6 sites in the USA

Main finding

As reported in the abstract

Antibiotic clearance due to the Seraph 100 device (CLSeraph)

Mean CLSeraph for the antibiotics investigated ranged from

−0.57 to 0.47 L/h

Original publication

Seraph 100 Microbind Affinity Blood Filter Does Not Clear Antibiotics: An Analysis of Antibiotic Concentration Data from PURIFY-OBS. Blood Purif. Published 13 January 2024. doi:10.1159/000531951