Peritoneal & home dialysis

Rapid point-of-care test for diagnosis of peritonitis in peritoneal dialysis patients

Perit Dial IntPublished Diagnostic accuracy study

Authors' conclusion

Periplex had high sensitivity, and positive and negative predictive values in the diagnosis of peritonitis can be considered as a screening tool for peritonitis. Given its high specificity and negative predictive value, it may also be used to document the resolution of peritonitis.

Study context

Population
Peritoneal dialysis (PD) patients suspected of having peritonitis, Singapore General Hospital, 2019–2022114 with peritonitis
Participants
120 patients
Outcome
Sensitivity and specificity of Periplex at presentation
Follow-up
At presentation
Methodological detailssingle-centre
  • Single-centre

Performance

As reported in the abstract

Diagnosis of peritonitis at presentation

Periplex

Sensitivity

100%

95% CI 100–100%

Specificity

50%

95% CI 41.1–59.0%

Positive predictive value

97.4%

Negative predictive value

100%

Periplex positive in all patients with peritonitis
114
Falsely positive in patients with non-infective eosinophilic peritonitis
3

Further findings

OutcomeValues
Resolution of infection during recovery from peritonitisPeriplex specificity 93.6% · NPV 98.7%

IL-6 vs MMP-8

MMP-8 more sensitive than IL-6 in detecting peritonitis

Type of PD solution

Periplex positive in all patients with peritonitis regardless of PD solution type

Original publication

Rapid point-of-care test for diagnosis of peritonitis in peritoneal dialysis patients. Perit Dial Int. Published 7 March 2024. doi:10.1177/08968608241234728

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.