Prediction of split renal function and obstruction with magnetic resonance urography in comparison with dynamic renal scintigraphy
Authors' conclusion
MRU could be an invaluable tool in the assessment of renal functions. Its diagnostic accuracy in detecting the level and extent of obstruction is comparable to that of MAG3 scintigraphy. Despite the lack of extensive comparative data, the potential benefits of MRU justify the need for further studies.
Study context
- Population
- Children who had previously undergone MAG3 scintigraphy, aged 1 to 216 months
- Participants
- 76 children
- Comparison
- Dynamic contrast-enhanced magnetic resonance urography (MRU) vs MAG3 scintigraphy
- Outcome
- Prediction of obstruction by MRU RTT (cut-off 11.5 min) and SRF agreement with scintigraphy
Methodological detailsretrospective comparison with MAG3 scintigraphy, CHOP-fMRU software, ROC analysis
- Retrospective comparison with MAG3 scintigraphy
- CHOP-fMRU software
- ROC analysis
Performance
As reported in the abstract
MRU
Sensitivity
94%
Specificity
86%
Positive predictive value
90%
Negative predictive value
88%
Accuracy 88%
- Positive correlation between split renal functions (SRFs) of right and left kidneys measured by MRU and scintigraphy (r > 0.9)
- P = 0.001
Key notes
- RTT cut-off 11.5 min
Original publication
Prediction of split renal function and obstruction with magnetic resonance urography in comparison with dynamic renal scintigraphy. Pediatr Nephrol. Published 17 February 2026. doi:10. 1007/ s00467-026-07211-y