PLA2R autoantibodies, a multifaceted biomarker in nephrotic syndrome and membranous nephropathy
Authors' conclusion
By evaluating the specificity and sensitivity of EUROIMMUN ELISA PLA2R-Ab and IF, we demonstrate that at ELISA levels ≥20 RU/mL, alongside thorough secondary screening, a kidney biopsy may be unnecessary.
Study context
- Population
- Patients with membranous nephropathy (MN) (PLA2R-related vs non-PLA2R MN)91 studies
- Comparison
- EUROIMMUN enzyme-linked immunosorbent assay (ELISA) vs EUROIMMUN immunofluorescence (IF) phospholipase A2 receptor antibody (PLA2R-Ab) test for distinguishing PLA2R-related MN from non-PLA2R MN
- Outcome
- Pooled diagnostic accuracy (sensitivity and specificity) of the PLA2R-Ab test
Methodological detailsQUADAS-2 risk of bias assessment, Cochrane-recommended methodology
- QUADAS-2 risk of bias assessment
- Cochrane-recommended methodology
Pooled result
As reported in the abstract
Pooled diagnostic accuracy for PLA2R-related MN
| Strategy | Sensitivity | Specificity |
|---|---|---|
| EUROIMMUN ELISA | 0.64 (95% CI 0.56-0.72) | 94.7% (95% CI 90.5-97.1%) |
| EUROIMMUN IF | 0.69 (95% CI 0.637-0.739) | 0.98 (95% CI 0.931-0.994) |
Sensitivity
EUROIMMUN ELISA0.64 (95% CI 0.56-0.72)
EUROIMMUN IF0.69 (95% CI 0.637-0.739)
Specificity
EUROIMMUN ELISA94.7% (95% CI 90.5-97.1%)
EUROIMMUN IF0.98 (95% CI 0.931-0.994)
Further findings
Risk of bias
- Higher risk of bias for studies evaluating IF vs ELISA
Timing of the index test
- No significant differences in pooled diagnostic accuracy by timing of the index test
Original publication
PLA2R autoantibodies, a multifaceted biomarker in nephrotic syndrome and membranous nephropathy. Nephrol Dial Transplant. Published 16 January 2025. doi:10. 1093/ ndt/ gfaf012