Performance of point-of-care creatinine testing in the estimation of glomerular filtration rate: comparison with laboratory creatinine and cystatin C-based measurements
Authors' conclusion
While point-of-care testing shows limitations for precise GFR estimation, it demonstrates utility for community-based CKD screening with high negative predictive value.
Study context
- Population
- Adults undergoing routine kidney function testing at Ouakam Military Hospital
- Participants
- 405 adults
- Exposure
- Point-of-care creatinine estimated GFR (eGFR) vs laboratory creatinine- and cystatin C-based eGFR
- Outcome
- Performance of point-of-care testing for estimating glomerular filtration rate (GFR) vs laboratory creatinine- and cystatin C-based eGFR
Methodological detailsanalytical, single-centre
- Analytical
- Single-centre
Main finding
As reported in the abstract
Mean bias
−31.09 mL/min/1.73 m²
Point-of-care creatinine vs Laboratory serum creatinine · eGFR; point-of-care overestimated creatinine, underestimating eGFR
Bias
−38.57 mL/min/1.73 m²
Point-of-care creatinine vs Cystatin C · significant negative bias in eGFR
- Correlation with laboratory serum creatinine (r)
- 0.48
- Correlation with laboratory creatinine eGFR (r)
- 0.60
- Correlation with cystatin C-based eGFR (r)
- 0.35
- P30 accuracy vs lab creatinine
- 32.6%
- P30 accuracy vs cystatin C
- 28.1%
Observational data: associations, not effects.
Further findings
Diagnostic performance of point-of-care creatinine for detecting chronic kidney disease (CKD) stage 3 and over (eGFR < 60 mL/min/1.73 m²)
- NPV, cystatin C as reference
- 91.3%
- NPV, serum creatinine as reference
- 90%
- High NPV