Performance Evaluation of an Automated Peritoneal Dialysis Solution Generation System in Patients Using Automated Peritoneal Dialysis
Authors' conclusion
This study demonstrated that the SGS can successfully generate PD solution in patients' homes, while meeting chemical composition and ISO microbiological standards. Lessons learned from this clinical trial will be useful in optimizing product development and future clinical trials.
Study context
- Population
- Prevalent adult patients with ESKD on maintenance peritoneal dialysis (PD) using automated PD (APD)14 completed
- Participants
- 22 patients
- Outcome
- Safety (microbiological testing) and efficacy (chemical composition) of the product water generated by the SGS
Methodological detailsmulticentre, home use, 12 US centres
- Multicentre
- Home use
- 12 US centres
Main finding
As reported in the abstract
Product water meeting ISO specifications (microbiological and chemical)
Samples meeting ISO specifications
all 64 samples
Safety
- Total adverse events (AEs)
- 34 in 12 patients
- Serious treatment-emergent AEs (culture negative peritonitis, bacterial peritonitis, AF)
- 3
Original publication
Performance Evaluation of an Automated Peritoneal Dialysis Solution Generation System in Patients Using Automated Peritoneal Dialysis. Kidney Int Rep. Published 18 March 2024. doi:10. 1016/ j. ekir. 2024. 03. 010