Authors' conclusion

This study demonstrated that the SGS can successfully generate PD solution in patients' homes, while meeting chemical composition and ISO microbiological standards. Lessons learned from this clinical trial will be useful in optimizing product development and future clinical trials.

Study context

Population
Prevalent adult patients with ESKD on maintenance peritoneal dialysis (PD) using automated PD (APD)14 completed
Participants
22 patients
Outcome
Safety (microbiological testing) and efficacy (chemical composition) of the product water generated by the SGS
Methodological detailsmulticentre, home use, 12 US centres
  • Multicentre
  • Home use
  • 12 US centres

Main finding

As reported in the abstract

Product water meeting ISO specifications (microbiological and chemical)

Samples meeting ISO specifications

all 64 samples

Safety

Total adverse events (AEs)
34 in 12 patients
Serious treatment-emergent AEs (culture negative peritonitis, bacterial peritonitis, AF)
3

Original publication

Performance Evaluation of an Automated Peritoneal Dialysis Solution Generation System in Patients Using Automated Peritoneal Dialysis. Kidney Int Rep. Published 18 March 2024. doi:10.1016/j.ekir.2024.03.010