Patterns in Heart Rate Variability during Hemodialysis
Authors' conclusion
HRV metrics did not significantly change over the course of hemodialysis sessions, suggesting poor autonomic function during dialysis. A novel fingertip sensor accurately measured time-domain (but not frequency-domain) HRV metrics when compared with a gold-standard ECG device.
Study context
- Population
- Patients treated with in-center HD
- Participants
- 22 patients
- Exposure
- Novel fingertip sensor vs gold-standard electrocardiogram (ECG) device
- Outcome
- Change in time-domain and frequency-domain HRV metrics over the course of HD sessions
- Follow-up
- Five time points during two consecutive dialysis sessions
Methodological detailsin-center hemodialysis, five time points during two consecutive dialysis sessions, linear mixed models, device comparison vs gold-standard ECG
- In-center hemodialysis
- Five time points during two consecutive dialysis sessions
- Linear mixed models
- Device comparison vs gold-standard ECG
Main finding
As reported in the abstract
No significant changes in heart rate variability (HRV) metrics over the dialysis session, ECG
P value for trend 0.11
No significant changes in HRV metrics over the dialysis session, CorSense
P value for trend 0.58
Observational data: associations, not effects.
Further findings
Agreement of CorSense with ECG, time-domain HRV metrics
- Correlation coefficient for SD of N-N interval
- 0.72
- Correlation coefficient for RMSSD
- 0.66
- Linear models suggested strong association for time-domain metrics
- Moderately correlated
Agreement of CorSense with ECG, frequency-domain HRV metrics
- Frequency-domain methods not correlated between the two methods
Original publication