Mid-Range Dialysate Flow and Adequacy of Dialysis
Authors' conclusion
The restriction of QD to 300 mL/min may compromise adequacy of dialysis in acute care setting. This may be mitigated by prolonging session times, albeit at the expense of higher resource utilization, and improving the QD:QB ratio. Further studies in a larger hospital cohort are necessary to elucidate these issues.
Study context
- Population
- Veterans with ESKD or dialysis-dependent AKI who completed ≥3 h of HD with pre-dialysis blood urea nitrogen (BUN) >20 mg/dL and available post-dialysis BUN, on a fully integrated HD machine with mid-range dialysate flow of 300 mL/min in the acute hospital setting59 dialysis sessions
- Exposure
- Sessions by dialyzer size and duration
- Outcome
- Efficiency of HD (URR and single-pool Kt/V) with mid-range dialysate flow rate of 300 mL/min
Methodological detailssingle-centre, February 2023 to June 2023
- Single-centre
- February 2023 to June 2023
Main finding
As reported in the abstract
Mean Kt/V
1.26 ± 0.04
Mean urea reduction ratio (URR)
60.8 ± 7.3%
Sessions achieving minimum adequate clearance per Kt/V criteria
50.6%
Sessions achieving minimum adequate clearance per URR criteria
32.5%
Observational data: associations, not effects.
Further findings
Sessions achieving minimum adequate Kt/V, by filter size and duration · subgroup analysis