Improvements in digital pathology equipment for renal biopsies: updating the standard model

J NephrolPublished Comparative instrument evaluation study

Authors' conclusion

The present report describes the experience of a hub nephropathology center adopting next generation digital pathology tools for the routine assessment of renal biopsies, highlighting the need for a complementary approach towards a philosophy of interoperability.

Study context

Population
Renal biopsy slides scanned in a nephropathology hub
Comparison
Hybrid MIDI II scanner vs Aperio scanner
Outcome
Scanning performance, light microscopy (default and custom settings)
Methodological detailssingle hub nephropathology centre, comparative quality control check of whole slide images
  • Single hub nephropathology centre
  • Comparative quality control check of whole slide images

Main finding

As reported in the abstract

  • No differences in number of out-of-focus slides per case (p = 0.75)
  • MIDI II vs Aperio: P < 0.001

Further findings

OutcomeP value
Scanning performance, immunofluorescenceLonger preparation time with the new scannerP = 0.001

Scanning performance, immunofluorescence

Comparable scanning times (p = 0.169) and final file size (p = 0.177)

Quality control of whole slide images

Differences in 3 quality features related to white background and blurriness (p < 0.001)

Diagnostic concordance

No major discordances in final diagnosis; minor disagreement
1 case (4%)

Original publication

Improvements in digital pathology equipment for renal biopsies: updating the standard model. J Nephrol. Published 14 February 2023. doi:10.1007/s40620-023-01568-1

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.