Higher blood flow rates in anticoagulation-free continuous renal replacement therapy (CRRT) improve circuit survival and clinical outcome
Authors' conclusion
A BFR of 250 ml/min significantly improved circuit survival in anticoagulation-free CRRT compared to 200 ml/min, likely mediated by TMP reduction. Moreover, patients receiving 250 ml/min BFR exhibited lower in-hospital mortality. Our results support adopting a higher BFR to optimize CRRT efficacy in anticoagulation-free settings. Larger-scale prospective trials are needed to validate the findings.
Study context
- Population
- Patients receiving anticoagulation-free CRRT in seven ICUs at West China Hospital of Sichuan University, September 2023–June 2024
- Participants
- 109 vs 132
- Exposure
- Blood flow rate (BFR) 250 ml/min vs 200 ml/min in anticoagulation-free CRRT
- 250 ml/minn = 109
- 200 ml/minn = 132reference
- Outcome
- Circuit lifespan: duration from CRRT initiation to termination due to clotting or other causes
Methodological detailsseven ICUs, single hospital, univariate and multivariate Cox proportional hazards regression, mediation analysis
- Seven ICUs, single hospital
- Univariate and multivariate Cox proportional hazards regression
- Mediation analysis
Main finding
As reported in the abstract
- Median circuit lifespan, clotted circuits: 250 vs 200 ml/min (hours)
- 31 vs. 15.5
- Median circuit lifespan, first circuit: 250 vs 200 ml/min (hours)
- 37 vs. 18
Key notes
- Median lifespans: all P < .05
- Statistically significant after confounder adjustment
Observational data: associations, not effects.
Secondary outcomes
Median circuit lifespan, clotted circuits
Median circuit lifespan, first circuit
Circuits achieving 12 h lifespan
Circuits achieving 24 h lifespan
Circuits achieving 48 h lifespan
Circuits achieving 72 h lifespan
In-hospital mortality
P = 0.005
- Median circuit lifespan, clotted circuits: all P < .05
- Median circuit lifespan, first circuit: all P < .05
- Circuits achieving 12 h lifespan: all P < .05
- Circuits achieving 24 h lifespan: all P < .05
- Circuits achieving 48 h lifespan: all P < .05
- Circuits achieving 72 h lifespan: all P < .05
Median circuit lifespan, overall
- All P < .05
Factors for circuit survival (multivariate)
- Protective: ST150 filter, exposure to non-CRRT anticoagulation agents, higher activated partial thromboplastin time
- Reduced circuit survival: higher substitute fluid rate, hyperlipidemia, hematocrit ≥0.30
Mediation by transmembrane pressure (TMP) increase
- Reduced TMP increase accounted for up to this share of the protective effect of higher BFR
- 60.32%