External Validation of a Limited Sampling Strategy for the Estimation of Mycophenolic Acid Exposure Between Different Assay Methods: Particle-enhanced Turbidimetric Inhibition Immunoassay (PETINIA) and HPLC Methods
Authors' conclusion
Our findings support the application of the MPA AUC0-12 LSS, originally developed using PETINIA measurements, to those obtained via HPLC-UV.
Study context
- Population
- Adult kidney transplant patients receiving mycophenolate mofetil32 development, 14 validation
- Comparison
- PETINIA vs HPLC-UV measurements for mycophenolic acid (MPA) AUC0-12 limited sampling strategy (LSS)
- Outcome
- Predictive performance of the LSS for MPA AUC0-12: MAE, RMSE and "good guess" (predicted AUC within observed AUC ± 15%)
Methodological detailsexternal validation, limited sampling strategy
- External validation
- Limited sampling strategy
Performance
As reported in the abstract
Predictive performance in validation sets
MAE
RMSE
Good guess
| Outcome | PETINIA | HPLC-UV |
|---|---|---|
| Coefficient of determination | 0.978 | 0.958 |
Key notes
- Comparable mean absolute error (MAE), root mean squared error (RMSE) and "good guess" outcomes between PETINIA and HPLC-UV
Further findings
Optimal LSS equation
- Three time point equation (0, 2 and 6 h) optimal for estimating MPA AUC0-12
Original publication
External Validation of a Limited Sampling Strategy for the Estimation of Mycophenolic Acid Exposure Between Different Assay Methods: PETINIA and HPLC Methods. Clin Transplant. Published 26 September 2024. doi:10. 1111/ ctr. 15471
Composed by AI from the published abstract and checked against it. How Briefs are made →