Clinical validation of the novel fluorescent glomerular filtration rate tracer agent relmapirazin (MB-102)
Authors' conclusion
Thus, relmapirazin may be used as a GFR tracer agent in humans.
Study context
- Population
- Adults with eGFR from normal to Stage 4 CKD
- Participants
- 120 adults3 clinical sites
- Comparison
- Relmapirazin vs iohexol plasma glomerular filtration rate (GFR)
- Outcome
- Agreement between plasma GFR measured by relmapirazin and by iohexol
- Follow-up
- 12 hours
Methodological detailsthree clinical sites, comparison to iohexol, two-compartment pharmacokinetic assessment
- Three clinical sites
- Comparison to iohexol
- Two-compartment pharmacokinetic assessment
Performance
As reported in the abstract
Linear regression correlation between the GFRs measured by the two tracers (r2)
0.99
Bland-Altman limits of agreement
−7.0 to +5.6 mL/min
Bland-Altman mean
−0.7 mL/min
Key notes
- Bland-Altman analysis confirmed agreement between the two measures of GFR
Further findings
GFR by CKD stage and race
- Relmapirazin GFR independent of stratification by CKD stage and by race
Urinary recovery of administered dose · 12 hours
- Percent of relmapirazin dose recovered in urine greater than or equal to that of iohexol
Safety
- No reported severe adverse events
Original publication
Clinical validation of the novel fluorescent glomerular filtration rate tracer agent relmapirazin (MB-102). Kidney Int. Published 2 July 2024. doi:10. 1016/ j. kint. 2024. 06. 012