CARPEDIEM® for continuous kidney replacement therapy in neonates and small infants: a French multicenter retrospective study
Authors' conclusion
These data confirm that the use of CARPEDIEM® is safe and effective in critically ill neonates and infants.
Study context
- Population
- Newborns and infants receiving continuous kidney replacement therapy (CKRT) with CARPEDIEM® in five French pediatric and neonatal ICUs19 newborns and 6 infants
- Participants
- 20 vs 5
- Exposure
- Continuous veno-venous hemofiltration (CVVH) vs Hemodialysis
- CVVHn = 20
- Hemodialysisn = 5
- Outcome
- Effectiveness, feasibility and technical parameters of CARPEDIEM® use
Methodological detailsmulticenter, five French pediatric and neonatal ICUs
- Multicenter
- Five French pediatric and neonatal ICUs
Main finding
As reported in the abstract
Effectiveness, feasibility and technical parameters of CARPEDIEM® use
| Group | Value | Effluent flow rate range | Net ultrafiltration (UF), median [IQR] (range) | Blood flow rate range | Dialysate flow rate, median [IQR] (range) | Session duration, median [IQR] (range) |
|---|---|---|---|---|---|---|
| Continuous veno-venous hemofiltrationCVVH | median 74 (IQR 43–99) mL/kg/h | 28-125 | 6 mL/kg/h [2-8] (1-12) | — not stated in the abstract | — not stated in the abstract | — not stated in the abstract |
| HD | median 6 (IQR 5–7) mL/kg/min | — not stated in the abstract | — not stated in the abstract | 4-7 | 600 mL/h [300-600] (120-600) | 8 h [6-12] (2-24) |
Value
Continuous veno-venous hemofiltrationCVVHmedian 74 (IQR 43–99) mL/kg/h
HDmedian 6 (IQR 5–7) mL/kg/min
Effluent flow rate range
Continuous veno-venous hemofiltrationCVVH28-125
HD— not stated in the abstract
Net ultrafiltration (UF), median [IQR] (range)
Continuous veno-venous hemofiltrationCVVH6 mL/kg/h [2-8] (1-12)
HD— not stated in the abstract
Blood flow rate range
Continuous veno-venous hemofiltrationCVVH— not stated in the abstract
HD4-7
Dialysate flow rate, median [IQR] (range)
Continuous veno-venous hemofiltrationCVVH— not stated in the abstract
HD600 mL/h [300-600] (120-600)
Session duration, median [IQR] (range)
Continuous veno-venous hemofiltrationCVVH— not stated in the abstract
HD8 h [6-12] (2-24)
- Age at initiation, median [IQR] (range), days
- 4 [2-13] (1-1134)
- Body weight at initiation, median [IQR] (range), kg
- 3.3 [2.5-4] (1.3-11.1)
- Total sessions
- 131
- Total treatment hours
- 2125
- Treatment duration per patient, median [IQR] (range), h
- 42 [24-91] (8-557)
- Blood flow rate, median [IQR] (range), mL/kg/min
- 8 [6-9] (3-16)
Key notes
- Most infants required a catheter between 4.5 and 6.5 French
- CVVH value is effluent flow rate; HD value is blood flow rate
Observational data: associations, not effects.
Further findings
Survival to hospital discharge
- Alive at hospital discharge
- 8 (32%)
- All patients survived the sessions
Safety
Hemodynamic instability needing volume replacement
- Sessions
- 31 (23%)
Thrombocytopenia
- Sessions
- 29 (22%)
Hemorrhage
- No hemorrhage occurred
Original publication
CARPEDIEM® for continuous kidney replacement therapy in neonates and small infants: a French multicenter retrospective study. Pediatr Nephrol. Published 10 January 2023. doi:10. 1007/ s00467-022-05871-0
Composed by AI from the published abstract and checked against it. How Briefs are made →