Authors' conclusion

In children immediately following kidney transplantation, clinically significant disagreement was found between invasive and noninvasive BP measurements. Invasive BP values were significantly higher than those obtained by Doppler. Better agreement was found between NIBP and Doppler. These issues should be considered when interpreting BP measurements in this sensitive patient population.

Study context

Population
Children aged 18 years or younger admitted to the pediatric ICU immediately following kidney transplantation82 paired measurements
Participants
18 children
Exposure
Invasive blood pressure (BP) vs noninvasive BP (NIBP), with systolic Doppler BP as reference
Outcome
Agreement between invasive BP and NIBP, compared with systolic Doppler BP
Follow-up
Immediately following kidney transplantation
Methodological detailsobservational, single-centre, tertiary, university-affiliated medical center, paired simultaneous measurements, May 2019 to June 2021
  • Observational
  • Single-centre
  • Tertiary, university-affiliated medical center
  • Paired simultaneous measurements
  • May 2019 to June 2021

Main finding

As reported in the abstract

Systolic BP, invasive vs NIBP

P < 0.001

Substantial differences (≥ 20 mmHg)
23% (95% CI 15-34%)

Observational data: associations, not effects.

Further findings

OutcomeInvasive BPNIBPDoppler BPP value
Systolic BP, invasive vs DopplerSimilar disagreement150 ± 23 mmHg— not stated in the abstract137 ± 17 mmHgP < 0.001
Systolic BP, NIBP vs DopplerGood agreement— not stated in the abstract135 ± 17138 ± 18P = 0.27

— not stated in the abstract

Correlation of higher systolic invasive BP with difference to systolic Doppler BP

Spearman's ρ (moderate to strong correlation)
0.63, p < 0.001

Hypertension

Mean systolic NIBP percentile (above the 95th percentile)
96 ± 6%

Original publication

Blood pressure monitoring following kidney transplantation in children: a comparison of invasive and noninvasive measurements using Doppler as a benchmark technique. Pediatr Nephrol. Published 1 August 2022. doi:10.1007/s00467-022-05691-2

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.