Blood pressure monitoring following kidney transplantation in children: a comparison of invasive and noninvasive measurements using Doppler as a benchmark technique
Authors' conclusion
In children immediately following kidney transplantation, clinically significant disagreement was found between invasive and noninvasive BP measurements. Invasive BP values were significantly higher than those obtained by Doppler. Better agreement was found between NIBP and Doppler. These issues should be considered when interpreting BP measurements in this sensitive patient population.
Study context
- Population
- Children aged 18 years or younger admitted to the pediatric ICU immediately following kidney transplantation82 paired measurements
- Participants
- 18 children
- Exposure
- Invasive blood pressure (BP) vs noninvasive BP (NIBP), with systolic Doppler BP as reference
- Outcome
- Agreement between invasive BP and NIBP, compared with systolic Doppler BP
- Follow-up
- Immediately following kidney transplantation
Methodological detailsobservational, single-centre, tertiary, university-affiliated medical center, paired simultaneous measurements, May 2019 to June 2021
- Observational
- Single-centre
- Tertiary, university-affiliated medical center
- Paired simultaneous measurements
- May 2019 to June 2021
Main finding
As reported in the abstract
Systolic BP, invasive vs NIBP
P < 0.001
- Substantial differences (≥ 20 mmHg)
- 23% (95% CI 15-34%)
Key notes
- Systolic invasive BP significantly higher than NIBP
Observational data: associations, not effects.
Further findings
| Outcome | Invasive BP | NIBP | Doppler BP | P value |
|---|---|---|---|---|
| Systolic BP, invasive vs DopplerSimilar disagreement | 150 ± 23 mmHg | — not stated in the abstract | 137 ± 17 mmHg | P < 0.001 |
| Systolic BP, NIBP vs DopplerGood agreement | — not stated in the abstract | 135 ± 17 | 138 ± 18 | P = 0.27 |
— not stated in the abstract
Correlation of higher systolic invasive BP with difference to systolic Doppler BP
- Spearman's ρ (moderate to strong correlation)
- 0.63, p < 0.001
Hypertension
- Mean systolic NIBP percentile (above the 95th percentile)
- 96 ± 6%
Original publication
Blood pressure monitoring following kidney transplantation in children: a comparison of invasive and noninvasive measurements using Doppler as a benchmark technique. Pediatr Nephrol. Published 1 August 2022. doi:10. 1007/ s00467-022-05691-2
Composed by AI from the published abstract and checked against it. How Briefs are made →