Application of therapeutic ultrasonic waves across the dialyzer membrane: A pilot study on the impact on dialyzer clearance and safety
Authors' conclusion
Intradialytic clogging of dialyzer membranes is a significant problem that can cause dialysis inadequacy. Our study tackles this issue by introducing therapeutic ultrasonic waves to improve dialyzer clearance during hemodialysis sessions in patients.
Study context
- Population
- Stable maintenance HD patients12 males · 3 females
- Participants
- 15 patients
- Comparison
- HD session with therapeutic ultrasonic waves vs HD session without ultrasonic waves
- Outcome
- Decline in dialyzer clearance (Online Clearance Monitoring)
Methodological detailseach patient own control
- Each patient own control
Main finding
As reported in the abstract
Mean percentage decline in dialyzer clearance
Ultrasonic session
4.41%
mean · SD 5.3
Control session
12.69%
mean · SD 6.35
P < 0.001
Safety
- Mean differences of blood component parameters comparable between Ultrasonic and Control sessions
- Microscopic membrane analysis corroborated safety
Original publication
Application of therapeutic ultrasonic waves across the dialyzer membrane: A pilot study on the impact on dialyzer clearance and safety. Hemodial Int. Published 24 May 2024. doi:10. 1111/ hdi. 13161